Z-0157-2020Class II

Selectra ProM Analyzer, Catalog # 6003-435/UDI # 03661540600395 and Catalog # 6003-436/UDI # 03661540600401

VitalScientific / initiated 2012-01-10 / Terminated / root cause: design

Vital Scientific N.V. began a recall of Selectra ProM Analyzer, Catalog # 6003-435/UDI # 03661540600395 and Catalog # 6003-436/UDI # 03661540600401 Product Usage: :clinical chemistry system intended... on . FDA classified it as Class II and gives the reason as "The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0157-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
VitalScientific
Product
Selectra ProM Analyzer, Catalog # 6003-435/UDI # 03661540600395 and Catalog # 6003-436/UDI # 03661540600401 Product Usage: :clinical chemistry system intended for use in clinical laboratories.
Product code
JGS Electrode, Ion Specific, Sodium
Reason for recall
The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced.
Root cause tag
design
FDA root cause
Component design/selection
Action
The firm reached out to consignees via phone to inform them of the affected product.
Quantity in commerce
5 instruments
Distribution
US Nationwide in the states of NY, and UT
Product codes and lots
serial numbers: 11-7425; 11-7516; 11-7535; 11-7563; 11-7564.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0157-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0154-2020Class IITerminatedVoluntary: Firm initiated
Z-0155-2020Class IITerminatedVoluntary: Firm initiated
Z-0156-2020Class IITerminatedVoluntary: Firm initiated
Z-0157-2020Class IITerminatedVoluntary: Firm initiated
Z-0158-2020Class IITerminatedVoluntary: Firm initiated