Selectra ProS Analyzer, Catalog # 6003-600/UDI # 03661540600418
Vital Scientific N.V. began a recall of Selectra ProS Analyzer, Catalog # 6003-600/UDI # 03661540600418 Product Usage: :clinical chemistry system intended for use in clinical laboratories on . FDA classified it as Class II and gives the reason as "The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0158-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- VitalScientific
- Product
- Selectra ProS Analyzer, Catalog # 6003-600/UDI # 03661540600418 Product Usage: :clinical chemistry system intended for use in clinical laboratories.
- Product code
- JGS Electrode, Ion Specific, Sodium
- Reason for recall
- The action was initiated due power supply failure. During a power supply failure the instruments malfunctions and no patient results are produced.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- The firm reached out to consignees via phone to inform them of the affected product.
- Quantity in commerce
- 6 instruments
- Distribution
- US Nationwide in the states of NY, and UT
- Product codes and lots
- serial numbers: 11-9609; 11-9690; 11-9694; 11-9695; 12-9618; 12-9620.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0154-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0155-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0156-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0157-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0158-2020 | Class II | Terminated | Voluntary: Firm initiated |