Z-2556-2019Class II
Arrow 8.5 Fr. X 16 cm Quad-Lumen Central Venous Catheterization Kit
Arrow International LLC (subsidiary of Teleflex Incorporated) / initiated 2019-08-06 / Terminated / root cause: packaging
Arrow International Inc began a recall of Arrow 8.5 Fr. X 16 cm Quad-Lumen Central Venous Catheterization Kit Product Code:AK-42854-P1A - Product Usage: The Arrow Central Venous Catheters are intended... on . FDA classified it as Class II and gives the reason as "Products may contain the incorrect banner card within the kit". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-2556-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Arrow International LLC (subsidiary of Teleflex Incorporated)
- Product
- Arrow 8.5 Fr. X 16 cm Quad-Lumen Central Venous Catheterization Kit Product Code:AK-42854-P1A - Product Usage: The Arrow Central Venous Catheters are intended to permit venous access to the central circulation by way of the femoral, jugular or subclavian veins.
- Product code
- DYB Introducer, Catheter
- Reason for recall
- Products may contain the incorrect banner card within the kit
- Root cause tag
- packaging
- FDA root cause
- Packaging process control
- Action
- Arrow International issued notification dated 8/6/19 stating problem, health risk and action to take: If you have affected stock in inventory, immediately discontinue use and quarantine any products with the product codes and lot numbers. 2. Inspect affected products within your control to confirm the product code on the lidstock matches the product code on the banner card of the same kit. Products confirmed to have an incorrect banner card should be returned to Arrow International. Products which properly match should not be returned. 3. If you have product with an incorrect banner card, please complete the enclosed Acknowledgement Form and fax it to 1-855-419-8507 Attn: Customer Service or email to [email protected]. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of products. 4. If you have no affected stock, please complete the enclosed Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to [email protected]. Questions, feel free to contact your local sales representative or Customer Service at 1-866-396-2111.
- Quantity in commerce
- 231 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Lot Number: 13F18H0499
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2555-2019 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2556-2019 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
