510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2538-2019Class II

Legion(tm) Hemi Step Tibial Wedge Rk/hk, 5 Mm, Size 3-4 Lt.mdl/rt-Lat, Ref 71423432

Smith & Nephew, Inc. / initiated 2018-01-15 / Completed / root cause: manufacturing process

Smith & Nephew, Inc. began a recall of LEGION(TM) HEMI STEP TIBIAL WEDGE RK/HK, 5 MM, SIZE 3-4 LT.MDL/RT-LAT, REF 71423432 Product Usage: The constrained knee systems are designed for use in... on . FDA classified it as Class II and gives the reason as "The affected products were inadvertently packaged with an incorrect screw component". As of , the recall status on record is Completed, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2538-2019
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
Smith & Nephew, Inc.
Product
LEGION(TM) HEMI STEP TIBIAL WEDGE RK/HK, 5 MM, SIZE 3-4 LT.MDL/RT-LAT, REF 71423432 Product Usage: The constrained knee systems are designed for use in patients in primary and revision surgery, where the posterior cruciate ligament and one or both of the collateral ligaments (i.e., medial collateral and/or lateral collateral ligament) are absent or incompetent.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The affected products were inadvertently packaged with an incorrect screw component.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm sent an Urgent Medical Device Recall Notice letter dated January 15, 2018 to consignees via email. The letter identified the affected product, problem and actions to be taken. Consignees are asked to inspect inventory, locate and quarantine affected, unused devices. Affected product will be returned to Smith & Nephew. All consignees are asked to complete the Response Form, even if they do not have product to return. For questions or concerns regarding this recall contact [email protected].
Quantity in commerce
12 units
Distribution
Foreign countries of; France, Germany, Netherlands, Norway, Switzerland
Product codes and lots
13KSL0038C

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2538-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2538-2019Class IICompletedVoluntary: Firm initiated