Z-2409-2019Class II
Monaco Radiation Treatment Planning System (RTP) System
Elekta Limited / initiated 2019-08-23 / Terminated / root cause: design
Elekta, Inc. began a recall of Monaco Radiation Treatment Planning System (RTP) System on . FDA classified it as Class II and gives the reason as "Monaco is displaying the Anatomy and Beam shift direction incorrectly on the Monaco Scan and Setup Reference Report and is DICOM exporting incorrect (Anatomy/Beam) shift directions". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2409-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Monaco Radiation Treatment Planning System (RTP) System
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- Monaco is displaying the Anatomy and Beam shift direction incorrectly on the Monaco Scan and Setup Reference Report and is DICOM exporting incorrect (Anatomy/Beam) shift directions.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Elekta Inc. notified customers on about 08/23/2019 via "IMPORTANT FIELD SAFETY NOTIFICATION" letter. The letter informed customers that the Monaco is displaying some incorrect shift information on the Scan and Setup Reference Report and is DICOM exporting incorrect shift information. Instructions included that all shifts should be verified before treatment, that a software fix is planned. Customer were also instructed to post the notice in a place accessible to all users, advise personnel working with the device of the issue, and to complete and return the Acknowledgement Form provided.
- Quantity in commerce
- 49 units
- Distribution
- Distribution within the US to Ohio and Indiana. International distribution to Germany, Greece, Sweden.
- Product codes and lots
- Software Build: 5.50.00 & 5.51.00; UDI GTIN: (01)00858164002244(10) 5.50.00, (01)00858164002268(10) 5.51.00
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2409-2019 | Class II | Terminated | Voluntary: Firm initiated |