Z-2409-2019Class II

Monaco Radiation Treatment Planning System (RTP) System

Elekta Limited / initiated 2019-08-23 / Terminated / root cause: design

Elekta, Inc. began a recall of Monaco Radiation Treatment Planning System (RTP) System on . FDA classified it as Class II and gives the reason as "Monaco is displaying the Anatomy and Beam shift direction incorrectly on the Monaco Scan and Setup Reference Report and is DICOM exporting incorrect (Anatomy/Beam) shift directions". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2409-2019
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Monaco Radiation Treatment Planning System (RTP) System
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
Monaco is displaying the Anatomy and Beam shift direction incorrectly on the Monaco Scan and Setup Reference Report and is DICOM exporting incorrect (Anatomy/Beam) shift directions.
Root cause tag
design
FDA root cause
Device Design
Action
Elekta Inc. notified customers on about 08/23/2019 via "IMPORTANT FIELD SAFETY NOTIFICATION" letter. The letter informed customers that the Monaco is displaying some incorrect shift information on the Scan and Setup Reference Report and is DICOM exporting incorrect shift information. Instructions included that all shifts should be verified before treatment, that a software fix is planned. Customer were also instructed to post the notice in a place accessible to all users, advise personnel working with the device of the issue, and to complete and return the Acknowledgement Form provided.
Quantity in commerce
49 units
Distribution
Distribution within the US to Ohio and Indiana. International distribution to Germany, Greece, Sweden.
Product codes and lots
Software Build: 5.50.00 & 5.51.00; UDI GTIN: (01)00858164002244(10) 5.50.00, (01)00858164002268(10) 5.51.00

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2409-2019
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2409-2019Class IITerminatedVoluntary: Firm initiated