Z-0288-2020Class II

(1) Argon Medical Devices Double Female LL Adapter, Special Adapter to Convert a Male Fitting to Female LL...

Argon Medical Devices, Inc / initiated 2018-10-16 / Terminated / root cause: manufacturing process

Argon Medical Devices, Inc began a recall of (1) Argon Medical Devices Double Female LL Adapter, Special Adapter to Convert a Male Fitting to Female LL, REF 040183000A, 25/box, Sterile, Rx; and (2) Argon... on . FDA classified it as Class II and gives the reason as "The seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0288-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Argon Medical Devices, Inc
Product
(1) Argon Medical Devices Double Female LL Adapter, Special Adapter to Convert a Male Fitting to Female LL, REF 040183000A, 25/box, Sterile, Rx; and (2) Argon Medical Devices Double Male LL Adapter, Special Adapter to Convert a Female Fitting to Male LL, REF 040184000A, 25/box, Sterile, Rx. The firm name on the labels is Argon Medical Devices, Inc., Athens, TX.
Product code
DRS Transducer, Blood-Pressure, Extravascular
Reason for recall
The seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
The recalling firm began issuing letters dated 10/16/2018 on 10/16/2018 via registered mail identifying the issue and the action to be taken by the customer.
Quantity in commerce
Total number in recall for all products in dist: 49,130 units
Distribution
Distribution was nationwide. There was military/government distribution. Foreign distribution was made to Canada, Australia, the Netherlands, and Spain.
Product codes and lots
(1) Double Female LL Adapter-special adapter to convert male fitting to a female luer lock - Lot numbers 11228606, 11227966, 11226655; and  (2) Double Male LL Adapter-special adapter to convert a female fitting to a male luer lock - Lot numbers 11227508, 11229885, 11230268, and 11231393.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0288-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0288-2020Class IITerminatedVoluntary: Firm initiated
Z-0289-2020Class IITerminatedVoluntary: Firm initiated
Z-0290-2020Class IITerminatedVoluntary: Firm initiated
Z-0291-2020Class IITerminatedVoluntary: Firm initiated
Z-0292-2020Class IITerminatedVoluntary: Firm initiated
Z-0293-2020Class IITerminatedVoluntary: Firm initiated
Z-0294-2020Class IITerminatedVoluntary: Firm initiated
Z-0295-2020Class IITerminatedVoluntary: Firm initiated
Z-0296-2020Class IITerminatedVoluntary: Firm initiated