Z-0293-2020Class II

(1) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Small Straight, REF 046-1-200, 2...

Argon Medical Devices, Inc / initiated 2018-10-16 / Terminated / root cause: manufacturing process

Argon Medical Devices, Inc began a recall of (1) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Small Straight, REF 046-1-200, 2 clamps/package, 5 packages/box, 10 boxes/case, Sterile, Rx... on . FDA classified it as Class II and gives the reason as "The seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0293-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Argon Medical Devices, Inc
Product
(1) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Small Straight, REF 046-1-200, 2 clamps/package, 5 packages/box, 10 boxes/case, Sterile, Rx; (2) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Medium Straight, REF 046-1-300, 2 clamps per package, 5 packages/ shelf box, 10 shelf boxes/case (for a total of 50/2-clamp packages/case), Sterile, Rx; (3) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Medium Angled, REF 046-1-301, 2 clamps per package, 6 packages/box, 50 boxes/case, Sterile, Rx; and (4) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Large Angled, REF 046-1-401, 2 clamps per package, 5 packages/box, 50 boxes/case. The firm name on the label is Argon Medical Devices, Inc., Athens, TX.
Product code
DRS Transducer, Blood-Pressure, Extravascular
Reason for recall
The seal formed on the edge of the pouch may be incomplete, which could result in a loss of sterility to the product.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
The recalling firm began issuing letters dated 10/16/2018 on 10/16/2018 via registered mail identifying the issue and the action to be taken by the customer.
Quantity in commerce
Total number in recall for all products in dist: 49,130 units
Distribution
Distribution was nationwide. There was military/government distribution. Foreign distribution was made to Canada, Australia, the Netherlands, and Spain.
Product codes and lots
(1) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Small Straight, REF 046-1-200 - Lot number 11230432;  (2) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Medium Straight, REF 046-1-300 - Lot numbers 11226690 and 11231581  (3) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Medium Angled, REF 046-1-301 - Lot number 11227669;  (4) Argon Medical Devices Vascuclamp, Radiopaque Vascular Clamp, Large Angled, REF 046-1-401 - Lot numbers 11228371 and 11230677.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0293-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0288-2020Class IITerminatedVoluntary: Firm initiated
Z-0289-2020Class IITerminatedVoluntary: Firm initiated
Z-0290-2020Class IITerminatedVoluntary: Firm initiated
Z-0291-2020Class IITerminatedVoluntary: Firm initiated
Z-0292-2020Class IITerminatedVoluntary: Firm initiated
Z-0293-2020Class IITerminatedVoluntary: Firm initiated
Z-0294-2020Class IITerminatedVoluntary: Firm initiated
Z-0295-2020Class IITerminatedVoluntary: Firm initiated
Z-0296-2020Class IITerminatedVoluntary: Firm initiated