AssayTip/AssayCup Tray-used on the cobas e 801 module which is a part of the cobas c 8000 MODULAR Analyzer...
Roche Diagnostics Operations, Inc. began a recall of AssayTip/AssayCup Tray-used on the cobas e 801 module which is a part of the cobas c 8000 MODULAR Analyzer Series (IVD for in vitro determination of analytes... on . FDA classified it as Class II and gives the reason as "AssayTips part of the AssayTip/AssayCup Tray Tips used on the cobas e 801 module- may have an Abnormal Internal Structure and lead to Incorrect patient results". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0044-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Roche Diagnostics GmbH
- Product
- AssayTip/AssayCup Tray-used on the cobas e 801 module which is a part of the cobas c 8000 MODULAR Analyzer Series (IVD for in vitro determination of analytes in human body fluids) Catalog Number: 5694302001
- Product code
- JLW Radioimmunoassay, Thyroid-Stimulating Hormone
- Reason for recall
- AssayTips part of the AssayTip/AssayCup Tray Tips used on the cobas e 801 module- may have an Abnormal Internal Structure and lead to Incorrect patient results
- Root cause tag
- packaging
- FDA root cause
- Process change control
- Action
- Roche Diagnostics issued Urgent Medical Notification letter to Distributor (TP-00744) and end users( TP-00741) on 8/13/19. Letter identifies problem, health risk and Actions required per the UMDC: " Check to see if you have any of the affected lot number(s) of AssayTips/AssayCup Trays on your cobas e 801 module or in your laboratory stock. (See pictures for lot number locations on packaging.). " Stop using the affected lot number(s) of AssayTips/AssayCup Trays immediately. " Discard the affected lot number(s) of AssayTips/AssayCup Trays according to your local waste management guidelines. To receive replacement product, fill out the attached fax form and fax or email it according to the instructions on the form. Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions
- Quantity in commerce
- 935 units
- Distribution
- Nationwide
- Product codes and lots
- Lot Numbers: 18636170 18665170 18667170 18668170 18215274 18245274 18250274 18251274 18252274 18253274 18265274 18271274 19209274
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0044-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY