Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0044-2020Class II

AssayTip/AssayCup Tray-used on the cobas e 801 module which is a part of the cobas c 8000 MODULAR Analyzer...

Roche Diagnostics GmbH / initiated 2019-08-13 / Terminated / root cause: packaging

Roche Diagnostics Operations, Inc. began a recall of AssayTip/AssayCup Tray-used on the cobas e 801 module which is a part of the cobas c 8000 MODULAR Analyzer Series (IVD for in vitro determination of analytes... on . FDA classified it as Class II and gives the reason as "AssayTips part of the AssayTip/AssayCup Tray Tips used on the cobas e 801 module- may have an Abnormal Internal Structure and lead to Incorrect patient results". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0044-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
AssayTip/AssayCup Tray-used on the cobas e 801 module which is a part of the cobas c 8000 MODULAR Analyzer Series (IVD for in vitro determination of analytes in human body fluids) Catalog Number: 5694302001
Product code
JLW Radioimmunoassay, Thyroid-Stimulating Hormone
Reason for recall
AssayTips part of the AssayTip/AssayCup Tray Tips used on the cobas e 801 module- may have an Abnormal Internal Structure and lead to Incorrect patient results
Root cause tag
packaging
FDA root cause
Process change control
Action
Roche Diagnostics issued Urgent Medical Notification letter to Distributor (TP-00744) and end users( TP-00741) on 8/13/19. Letter identifies problem, health risk and Actions required per the UMDC: " Check to see if you have any of the affected lot number(s) of AssayTips/AssayCup Trays on your cobas e 801 module or in your laboratory stock. (See pictures for lot number locations on packaging.). " Stop using the affected lot number(s) of AssayTips/AssayCup Trays immediately. " Discard the affected lot number(s) of AssayTips/AssayCup Trays according to your local waste management guidelines. To receive replacement product, fill out the attached fax form and fax or email it according to the instructions on the form. Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions
Quantity in commerce
935 units
Distribution
Nationwide
Product codes and lots
Lot Numbers: 18636170 18665170 18667170 18668170 18215274 18245274 18250274 18251274 18252274 18253274 18265274 18271274 19209274

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0044-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0044-2020Class IITerminatedVoluntary: Firm initiated