Philips HeartStart Infant/Child SMART Pads Cartridge, Model # M5072A, UDI:00884838023758
Philips North America LLC began a recall of Philips HeartStart Infant/Child SMART Pads Cartridge, Model # M5072A, UDI:00884838023758 on . FDA classified it as Class II and gives the reason as "The recall is being initiated due to incorrect age range labeling (8+ yrs) on the foil packaging for infant/child pad cartridges. The infant/child pad cartridge is designed for use with the defibrillator An Infant/Child...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2549-2019
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips HeartStart Infant/Child SMART Pads Cartridge, Model # M5072A, UDI:00884838023758
- Product code
- MKJ Automated External Defibrillators (Non-Wearable)
- Reason for recall
- The recall is being initiated due to incorrect age range labeling (8+ yrs) on the foil packaging for infant/child pad cartridges. The infant/child pad cartridge is designed for use with the defibrillator An Infant/Child pad cartridge should not be used on an adult because it delivers lower energy (50 J vs 150 J) than the Adult pads cartridge and may therefore not be effective on an adult.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling False and Misleading
- Action
- A labeling specification for the HS1 is available against which labeling can be checked prior to design transfer and/or localization. This activity has been moved under regulatory affairs for review and maintenance.
- Quantity in commerce
- 181,894 Infant/Child SMART Pads Pads
- Distribution
- US: AK, AL,AR,AZ,CA,CO,CT,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH,NJ,NM,NV,NY,OH,OK,OR,PA,PR,RI,SC,SD,TN,TX,UT,VA,VT,WA,WI,WV, WY, Argentina, Aruba, Australia, Austria, Belgium, Bolivia, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Denmark, Dominican Republic, Egypt, El Salvador, France, French Guiana, Germany, Greece, Guadeloupe, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Republic of, Malaysia, Malta, Martinique, Mauritius, Mexico, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Norway, Panama, Paraguay, Peru, Philippines, Poland, Puerto Rico, R¿union, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, United Kingdom, Uruguay, Viet Nam,
- Product codes and lots
- Model# M5072A; Lots #s Y070115-04 - Y011118-07;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2549-2019 | Class II | Terminated | Voluntary: Firm initiated |