Z-0350-2020Class II
Universal Sexual Health Testing Kit
Privapath Diagnostics Ltd / initiated 2019-07-01 / Terminated / root cause: manufacturing process
Privapath Diagnostics Ltd began a recall of Universal Sexual Health Testing Kit on . FDA classified it as Class II and gives the reason as "A supplier informed PrivaPath Diagnostics that there is a potential for damaged tubes which could result in blood exposure or erroneous results due to an inappropriate blood-to-additive ratio". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0350-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Privapath Diagnostics Ltd
- Product
- Universal Sexual Health Testing Kit
- Product code
- N/A
- Reason for recall
- A supplier informed PrivaPath Diagnostics that there is a potential for damaged tubes which could result in blood exposure or erroneous results due to an inappropriate blood-to-additive ratio.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- PrivaPath Diagnostics telephoned and or issued emails to all customers and stated the following: PrivaPath Diagnostics has not received any reports from any customers or laboratory partners regarding this issue and the telephone call/email is just a precautionary action. However, PrivaPath Diagnostics PrivaPath Diagnostics is working closely with the supplier to ensure that this issue does not recur. If contact has been unsuccessful the customer will be sent a second email. This action ensures that in the case of a potential erroneous result it is followed up upon either through a replacement kit with LetsGetChecked with an unaffected lot number or with another healthcare provider.
- Distribution
- Nationwide
- Product codes and lots
- lot numbers: 1189 1094 144 226 82
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0350-2020 | Class II | Terminated | Voluntary: Firm initiated |