Z-0362-2020Class II

TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580002 -

Tec Com GmbH / initiated 2018-08-14 / Terminated / root cause: manufacturing process

Tec Com Gmbh began a recall of TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580002 - Product Usage: The Thermo-Regulation System... on . FDA classified it as Class II and gives the reason as "The action is being initiated due to potential installation of faulty peltier-elements within the thermo-regulation system. Failure of the peltier-element would report a system failure and no longer be able to guarantee...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0362-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Tec Com GmbH
Product
TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580002 - Product Usage: The Thermo-Regulation System TECOTHERM NEO is designed for controlled cold & heat treatment procedures. By means of a mattress, cold and heat is provided to the patient, depending on the therapy objective.
Product code
DWJ System, Thermal Regulating
Reason for recall
The action is being initiated due to potential installation of faulty peltier-elements within the thermo-regulation system. Failure of the peltier-element would report a system failure and no longer be able to guarantee temperature control at full power.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Quantity in commerce
5 units
Distribution
US Nationwide distribution in the state of UT.
Product codes and lots
Serial Numbers: 2018/09/07, 2018/10/01, 2018/10/02, 2018/10/03, 2018/10/05

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0362-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0362-2020Class IITerminatedVoluntary: Firm initiated