Custom Procedure Kit, Custom Kit, REF/
Merit Medical Systems, Inc. began a recall of Custom Procedure Kit, Custom Kit, REF/Product Code: K12T-03688, single use, RX Only, Sterile EO, LATEX FREE, UDI: (01)00884450389409 on . FDA classified it as Class II and gives the reason as "Labeling error; It was identified that a Procedure Kit/Tray contained incorrect latex labeling information (i.e. indicated Latex Free)". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0026-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Merit Medical Systems, Inc.
- Product
- Custom Procedure Kit, Custom Kit, REF/Product Code: K12T-03688, single use, RX Only, Sterile EO, LATEX FREE, UDI: (01)00884450389409
- Product code
- OEQ Angiography/Angioplasty Kit
- Reason for recall
- Labeling error; It was identified that a Procedure Kit/Tray contained incorrect latex labeling information (i.e. indicated Latex Free)
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- On 10/26/2018, Merit customer service representative placed telephone calls to consignees to inform them of the latex contained in the kits/trays. On 10/29/2018, the firm sent an "Urgent Product recall Notice" to all affected consignees via FedEx. The recall notification informed customers of the following: 1. Please immediately determine if any of the devices identified in the attached Customer Response Form (CRF) are within your facility, quarantine them, and discontinue use and distribution. 2. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the CRF. Additional distribution details may be required by health authorities. 4. Please fill out, scan and email the completed CRF to [email protected] within 10 business days. 5. Please immediately return all affected lots in your possession to Merit, per the instructions found in the CRF. 6. If you have any questions concerning this communication, please contact Merit Customer service at (801) 208-4381 or by email at [email protected].
- Quantity in commerce
- 88 Kits
- Distribution
- U.S.: VA, IL
- Product codes and lots
- Catalog #K12T-03688H - Lot #T1401193 and #T1419935
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0026-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0027-2020 | Class II | Terminated | Voluntary: Firm initiated |