Z-0109-2020Class II

cobas c 701 and 702 Analyzer/Module-RF interference claims for the sTfR assay Catalog Number: 05950821190

Roche Diagnostics GmbH / initiated 2019-08-28 / Terminated

Roche Diagnostics Operations, Inc. began a recall of cobas c 701 and 702 Analyzer/Module-RF interference claims for the sTfR assay Catalog Number: 05950821190 on . FDA classified it as Class II and gives the reason as "Tina-quant Soluble Transferrin Receptor (STFR) Assays Updated Claims for Rheumatoid Factors Interference". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0109-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
cobas c 701 and 702 Analyzer/Module-RF interference claims for the sTfR assay Catalog Number: 05950821190
Product code
DDG Transferrin, Antigen, Antiserum, Control
Reason for recall
Tina-quant Soluble Transferrin Receptor (STFR) Assays Updated Claims for Rheumatoid Factors Interference
FDA root cause
Under Investigation by firm
Action
Roche issued notification to consignees by Urgent Medical Device Correction (TP-00754) sent via UPS (signature required) on 8/28/19. The UMDC will also be available on diagnostics.roche.com. The letter identifies the problem, risk to health and action to take: Be aware of new interference claims that are listed in this UMDC and that the rhEPO Therapy Monitoring Statement will be removed from the Method Sheets. The UMDC will serve as intermediate labeling until the updated method sheets are available. Complete all sections of the enclosed fax form. Contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions if you have questions
Quantity in commerce
1,086 units
Distribution
Nationwide
Product codes and lots
All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0109-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0108-2020Class IITerminatedVoluntary: Firm initiated
Z-0109-2020Class IITerminatedVoluntary: Firm initiated