YUKON Polyaxial Screws (3.5 X 12 mm, Owning Site: Leesburg USA 3.5 X 14 mm, 3.5 X 32 mm, 3.5 X 34 mm, 3.5 X...
K2M, Inc began a recall of YUKON Polyaxial Screws (3.5 X 12 mm, Owning Site: Leesburg USA 3.5 X 14 mm, 3.5 X 32 mm, 3.5 X 34 mm, 3.5 X 38 mm, 3.5 X 40 mm, 4.0 X 16 mm, 4.0 X 18 mm, 4.0 X... on . FDA classified it as Class II and gives the reason as "Screws had a manufacturing error that could prevent the screw from mating with the screw inserter or set screw as intended. This event was reported to FDA as part of Stryker's retrospective review of K2M product field...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0110-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- K2M, Inc
- Product
- YUKON Polyaxial Screws (3.5 X 12 mm, Owning Site: Leesburg USA 3.5 X 14 mm, 3.5 X 32 mm, 3.5 X 34 mm, 3.5 X 38 mm, 3.5 X 40 mm, 4.0 X 16 mm, 4.0 X 18 mm, 4.0 X 20 mm, 4.0 X 22 mm, 4.0 X 24 mm, 4.0 X 26 mm, 4.0 X 28 mm, 4.0 X 30 mm, 5.0 X 22 mm, 5.0 X 28 mm, 5.0 X 30 mm) Catalog Numbers 7601-03512, 7601-03514, 7601-03532, 7601-03534, 7601-03538L, 7601-03540L, 7601-04016L, 7601-04018L, 7601-04020L, 7601-04022L, 7601-04024L, 7601-04026L, 7601-04028L, 7601-04030, 7601-05022, 7601-05028, 7601-05030
- Product code
- NKG Posterior Cervical Screw System
- Reason for recall
- Screws had a manufacturing error that could prevent the screw from mating with the screw inserter or set screw as intended. This event was reported to FDA as part of Stryker's retrospective review of K2M product field actions; the action was completed in 2018.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- K2M notified customers with a Recall Notification Letter dated July 24, 2018. The letter instructed customers to locate and segregate recalled product, complete a tracking/verification form, and return the product to K2M.
- Quantity in commerce
- 364
- Distribution
- Distributed to accounts in AL, CA, CO, FL, IL, KY, LA, MD, MI, MN, MS, NC, NJ, NY, OH, OK, PA, TN, TX, and WA.. Foreign distribution to Australia.
- Product codes and lots
- Lots FUVD, FUVE, FUVG, FUWF, FUWK, FUWL, FUWU, FUWV, FUWW, FUXK, FUXL,FUXM, FUXN, FUXP, FUXR, FUXT, FUXU, FUXV
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|