Z-0031-2020Class II
TxRx 15Ch Knee Coil 3T QED part number:Q7000008 Siemens Model # 10185460
Quality Electrodynamics / initiated 2019-08-20 / Terminated / root cause: manufacturing process
Quality Electrodynamics LLC began a recall of TxRx 15Ch Knee Coil 3T QED part number:Q7000008 Siemens Model # 10185460 on . FDA classified it as Class II and gives the reason as "Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0031-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Quality Electrodynamics
- Product
- TxRx 15Ch Knee Coil 3T QED part number:Q7000008 Siemens Model # 10185460
- Product code
- MOS Coil, Magnetic Resonance, Specialty
- Reason for recall
- Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm sent a notification letter to their customers who that have received the affected device. Customers were instructed to. 1. Immediately examine your inventory and quarantine product subject to recall. 2. if you have further distributed this product to MRI end users, please identify your affected customers and notify them. 3. Please ensure they understand to stop use of the coil for scanning and to notify you immediately.
- Quantity in commerce
- 2 units
- Distribution
- Worldwide distribution - US in the states: NC OUS: Germany
- Product codes and lots
- Serial # 1459 1460
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0031-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0032-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0033-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0034-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0035-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0036-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0037-2020 | Class II | Terminated | Voluntary: Firm initiated |