Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0034-2020Class II

TxRx 15CH Knee Coil 1.5T QED part number: Q7000050 Siemens Model # 10606524

Quality Electrodynamics / initiated 2019-08-20 / Terminated / root cause: manufacturing process

Quality Electrodynamics LLC began a recall of TxRx 15CH Knee Coil 1.5T QED part number: Q7000050 Siemens Model # 10606524 on . FDA classified it as Class II and gives the reason as "Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0034-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Quality Electrodynamics
Product
TxRx 15CH Knee Coil 1.5T QED part number: Q7000050 Siemens Model # 10606524
Product code
MOS Coil, Magnetic Resonance, Specialty
Reason for recall
Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm sent a notification letter to their customers who that have received the affected device. Customers were instructed to. 1. Immediately examine your inventory and quarantine product subject to recall. 2. if you have further distributed this product to MRI end users, please identify your affected customers and notify them. 3. Please ensure they understand to stop use of the coil for scanning and to notify you immediately.
Quantity in commerce
3 units
Distribution
Worldwide distribution - US in the states: NC OUS: Germany
Product codes and lots
Serial # 1433 1702 1440

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0034-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0031-2020Class IITerminatedVoluntary: Firm initiated
Z-0032-2020Class IITerminatedVoluntary: Firm initiated
Z-0033-2020Class IITerminatedVoluntary: Firm initiated
Z-0034-2020Class IITerminatedVoluntary: Firm initiated
Z-0035-2020Class IITerminatedVoluntary: Firm initiated
Z-0036-2020Class IITerminatedVoluntary: Firm initiated
Z-0037-2020Class IITerminatedVoluntary: Firm initiated