Z-0036-2020Class II
TxRx Knee 15 Coil 1.5T QED part number: Q7000056 Siemens Model # 10606828
Quality Electrodynamics / initiated 2019-08-20 / Terminated / root cause: manufacturing process
Quality Electrodynamics LLC began a recall of TxRx Knee 15 Coil 1.5T QED part number: Q7000056 Siemens Model # 10606828 on . FDA classified it as Class II and gives the reason as "Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0036-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Quality Electrodynamics
- Product
- TxRx Knee 15 Coil 1.5T QED part number: Q7000056 Siemens Model # 10606828
- Product code
- MOS Coil, Magnetic Resonance, Specialty
- Reason for recall
- Unintended ferrous material may be present within the pushbutton assembly of Siemens Knee and Siemens Extremity Coils.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The firm sent a notification letter to their customers who that have received the affected device. Customers were instructed to. 1. Immediately examine your inventory and quarantine product subject to recall. 2. if you have further distributed this product to MRI end users, please identify your affected customers and notify them. 3. Please ensure they understand to stop use of the coil for scanning and to notify you immediately.
- Quantity in commerce
- 36 units
- Distribution
- Worldwide distribution - US in the states: NC OUS: Germany
- Product codes and lots
- Serial # 1694 1841 1887 1933 1998 3195 3291 3292 3293 3294 3298 3299 3300 3301 3302 3303 3304 3305 3306 3307 3308 3309 3310 3311 3312 3313 3314 3315 3316 3317 3318 ¿1137 ¿1361 ¿1378 ¿1628 ¿2447
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0031-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0032-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0033-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0034-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0035-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0036-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0037-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.