510(K) CLEARANCES 67 ▼8RECALLS POSTED 26CLASS I 2PMA DECISIONS 40 ▼14WEEK ENDING / FDA DATA AS OF
Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0052-2020Class II

K2M Special Connector Instruments (Special Connector, Special Adaptor, and Quick Adaptor)

K2M, Inc / initiated 2016-12-02 / Terminated / root cause: manufacturing process

K2M, Inc began a recall of K2M Special Connector Instruments (Special Connector, Special Adaptor, and Quick Adaptor) Product Usage: Special Connector Instruments include an additional... on . FDA classified it as Class II and gives the reason as "Instruments were developed without adequate consideration given to the need for a new regulatory filing. This event was reported to FDA as part of Stryker's retrospective review of K2M product field actions; recall...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0052-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
K2M, Inc
Product
K2M Special Connector Instruments (Special Connector, Special Adaptor, and Quick Adaptor) Product Usage: Special Connector Instruments include an additional feature on their shafts that enable surgeon users to combine them with non-K2M navigation systems as part of their standard surgical technique.
Product code
OLO Orthopedic Stereotaxic Instrument
Reason for recall
Instruments were developed without adequate consideration given to the need for a new regulatory filing. This event was reported to FDA as part of Stryker's retrospective review of K2M product field actions; recall actions were completed in 2017.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
K2M notified customers with a Recall Notification Letter dated December 2, 2016. The letter instructed customers to locate and segregate recalled product, complete a tracking/verification form, and return the product to K2M. The product was reintroduced to the market after K2M obtained new 510(k) clearance for the product. for question call at 571-919-2000
Quantity in commerce
503
Distribution
US Nationwide Distribution in the states of AL CA CO CT FL FL IL IN KS MA MI MN PA TN TX UT VA
Product codes and lots
All product lots in distribution as of December 1, 2016, are in scope.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0052-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0052-2020Class IITerminatedVoluntary: Firm initiated