Z-0522-2020Class II
HENNESSY LEFT CONST GLENOID HA, Item No. PM555360 UDI (01)00880304819597(17)260425(10)664030
Zimmer GmbH / initiated 2019-08-21 / Terminated / root cause: manufacturing process
Zimmer Biomet, Inc. began a recall of HENNESSY LEFT CONST GLENOID HA, Item No. PM555360 UDI (01)00880304819597(17)260425(10)664030 Product Usage: Constrained Shoulder on . FDA classified it as Class II and gives the reason as "Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0522-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- HENNESSY LEFT CONST GLENOID HA, Item No. PM555360 UDI (01)00880304819597(17)260425(10)664030 Product Usage: Constrained Shoulder.
- Product code
- NA
- Reason for recall
- Elevated levels of bacterial endotoxin and residual debris remain on the devices due to cleaning issue.
- Root cause tag
- manufacturing process
- FDA root cause
- Environmental control
- Action
- On September 11, 2019, the firm began notifying distributors and customers of the recall via an Urgent Medical Device Recall letter. The letter informed consignees of the product issue. Customers were asked to assist their Zimmer Biomet sales representative and quarantine all identified product. The sales representative will remove the product from your facility. If any products were further distributed, please forward the notification to responsible parties. If you have further questions or concerns, please call customer service at 574-371-3071 between 8:00 am and 5:00 pm ET, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule.
- Quantity in commerce
- 219988 units (total)
- Distribution
- US Nationwide distribution and countries of Argentina, Australia, Brazil, Canada, Chile, Colombia, Costa Rica, Ecuador, El Salvador, Italy, Netherlands, India, Japan, Korea, Malaysia, Mexico, New Zealand, Panama, Singapore, Taiwan, Thailand, Trinidad and Tobago, and Venezuela.
- Product codes and lots
- 664030
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0519-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0520-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0521-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0522-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0523-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0524-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0525-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0526-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0527-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0528-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0529-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0530-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0531-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0532-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0533-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.