Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01...
Abbott Laboratories began a recall of Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01. The firm name on the label is Abbott... on . FDA classified it as Class II and gives the reason as "The diluent tubing used on the instrument appears cloudy or to have a powder/film on the inner and outer surfaces which may result in falsely elevated platelet results". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0173-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abbott Medical
- Product
- Emerald Diluent Tubing, List number 09H50-01, accompanying CELL-DYN Emerald instrument, List number 09H39-01. The firm name on the label is Abbott Laboratories, Abbott Park, IL.
- Product code
- GKZ Counter, Differential Cell
- Reason for recall
- The diluent tubing used on the instrument appears cloudy or to have a powder/film on the inner and outer surfaces which may result in falsely elevated platelet results.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The recalling firm issued letters on 5/9/2019 to the consignees who had the installed instruments with the affected serial numbers. The letters were issued via FedEx priority overnight express.
- Quantity in commerce
- 22 analyzers with tubing
- Distribution
- Distribution was made to GA, LA, MA, NC, NY, SC, and TX. There was no military/government distribution. Foreign distribution was made to Australia, Germany, Greece, Honduras, Indonesia, Italy, and Serbia.
- Product codes and lots
- Tubing lot #2018-10-16; Instrument serial numbers 007689, 007690, 007691, 007692, 007705, 007706, 007707, 007708, 007709, 007728, 007694, 007695, 007697, 007766, 007767, 007768, 007765, 007710, 007699, 007700, 007701, and 007703.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0173-2020 | Class II | Terminated | Voluntary: Firm initiated |