Z-0247-2020Class II
TherMax Blood Warmer Unit
Baxter Healthcare Corporation / initiated 2019-09-19 / Terminated / root cause: design
Baxter Healthcare Corporation began a recall of TherMax Blood Warmer Unit on . FDA classified it as Class II and gives the reason as "TherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0247-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Baxter Healthcare Corporation
- Product
- TherMax Blood Warmer Unit
- Product code
- KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
- Reason for recall
- TherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Baxter Healthcare notified customers on about 09/19/2019, via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that specific TherMax Blood Warmers listed below may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds. The affected units may not have an adequate ground path, which could cause the units to become damaged if used with improperly grounded outlets. The damage could cause the TherMax Blood Warmer units to be nonfunctional during clinical use. Instructions included that customers could continue to safely use affected devices with a properly grounded outlet, that a local Baxter representative will contact each facility to arrange for the correction of affected devices, to complete and return the Baxter Customer Reply Form via fax to 224-270-5457 or scanning and e-mailing it to [email protected], and to notify customers if the product was further distributed. For general questions regarding this communication, contact Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday.
- Quantity in commerce
- 504 units total
- Distribution
- Nationwide distribution to Georgia and Maryland. International distribution to Belgium, Germany, Finland, France, Greece, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom, Australia, and New Zealand
- Product codes and lots
- Product Code: 955630, UDI: 00085412639505, Serial Numbers: 101332, 101335, 101358, 101369, 101402, 101403, 101404, 101406, 101415, 101416, 101429, 101430, 101437, 101439, 101440, 101441, 101443. Product Code: 955515, Serial Numbers 100261, 100263, 100280, 100281, 100282, 100283, 100284, 100285, 100286, 100287, 100288, 100289, 100290, 100291, 100292, 100294, 100296, 100298, 100299, 100306, 100307, 100308, 100309, 100310, 100312, 100313, 100314, 100315, 100321, 100322, 100323, 100324, 100325, 100326, 100327, 100328, 100330, 100341, 100342, 100343, 100344, 100345, 100346, 100347, 100348, 100349, 100356, 100357, 100358, 100359, 100360, 100361, 100362, 100363, 100364, 100365, 100366, 100367, 100368, 100369, 100370, 100391, 100392, 100393, 100395, 100396, 100397, 100398, 100399, 100400, 100401, 100402, 100403, 100404, 100405, 100416, 100417, 100418, 100419, 100420, 100421, 100422, 100423, 100424, 100427, 100428, 100429, 100446, 100450, 100453, 100454, 100455, 100456, 100458, 100459, 100460, 100486, 100487, 100489, 100490, 100496, 100497, 100498, 100499, 100500, 100501, 100502, 100503, 100504, 100505, 100532, 100533, 100535, 100537, 100538, 100539, 100540, 100541, 100545, 100548, 100550, 100551, 100552, 100555, 100581, 100582, 100583, 100584, 100585, 100586, 100587, 100590, 100591, 100595, 100596, 100599, 100601, 100602, 100603, 100605, 100631, 100632, 100633, 100639, 100640, 100641, 100642, 100643, 100644, 100645, 100646, 100649, 100650, 100652, 100653, 100654, 100655, 100756, 100757, 100758, 100759, 100761, 100762, 100763, 100764, 100765, 100766, 100767, 100768, 100769, 100770, 100771, 100772, 100773, 100774, 100775, 100776, 100777, 100778, 100779, 100780, 100781, 100782, 100783, 100784, 100785, 100786, 100787, 100788, 100789, 100790, 100791, 100793, 100794, 100795, 100796, 100798, 100799, 100800, 100801, 100802, 100803, 100804, 100805, 100806, 100807, 100808, 100810, 100811, 100812, 100813, 100814, 100815, 100816, 100817, 100818, 100819, 100820, 100821, 100822, 100823, 100825, 100826, 100827, 100829, 100830, 100856, 100857, 100858, 100859, 100860, 100861, 100863, 100864, 100865, 100868, 100869, 100871, 100872, 100874, 100875, 100878, 100879, 100880, 100881, 100882, 100883, 100884, 100885, 100886, 100887, 100888, 100889, 100890, 100891, 100892, 100893, 100894, 100895, 100896, 100897, 100898, 100899, 100900, 100931, 100932, 100933, 100934, 100936, 100938, 100939, 100940, 100941, 100942, 100943, 100945, 100946, 100947, 100948, 100949, 100950, 100952, 100954, 100955, 100981, 100983, 100984, 100985, 100986, 100987, 100988, 100989, 100990, 100991, 100992, 100997, 100998, 100999, 101000, 101001, 101002, 101003, 101004, 101005, 101006, 101007, 101008, 101009, 101010, 101011, 101012, 101013, 101014, 101015, 101016, 101017, 101018, 101019, 101020, 101021, 101022, 101024, 101026, 101027, 101030, 101081, 101082, 101083, 101084, 101085, 101086, 101088, 101089, 101090, 101091, 101092, 101094, 101095, 101096, 101097, 101098, 101099, 101101, 101103, 101104, 101105, 101106, 101107, 101108, 101109, 101110, 101111, 101112, 101115, 101121, 101122, 101123, 101124, 101126, 101127, 101172, 101173, 101174, 101175, 101176, 101177, 101179, 101180, 101181, 101182, 101183, 101184, 101185, 101186, 101187, 101188, 101189, 101190, 101192, 101229, 101230, 101231, 101233, 101235, 101236, 101237, 101241, 101246, 101247, 101248, 101249, 101250, 101251, 101252, 101253, 101254, 101255, 101256, 101257, 101258, 101259, 101271, 101272, 101273, 101274, 101276, 101277, 101278, 101458, 101460, 101469, 101470, 101471, 101472, 101473, 101474, 101475, 101476, 101477, 101478, 101479, 101480, 101481, 101482, 101483, 101484, 101485, 101486, 101487, 101488, 101489, 101490, 101491, 101492, 101508, 101509, 101510, 101511, 101512, 101513, 101514, 101515, 101516, 101517, 101519, 101520, 101521, 101522, 101538, 101539, 101540, 101541, 101542, 101543, 101544, 101545, 101546, 101547, 101548, 101549, 101550, 101551, 101552, 101553, 101554, 101555, 101556, 101557, 101558, 101559, 101560, 101561, 101562, 101593, 101594, 101595, 101596, 101597, 101598, 101599, 101600, 101601, 101602, 101603, 101604, 101605, 101607.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0247-2020 | Class II | Terminated | Voluntary: Firm initiated |