TriMed Wrist Hook Dorsal Impactor, REF IMPCT-WHD; and TriMed Wrist Hook Volar Impactor, REF IMPCT-WHV, both...
Trimed Inc / initiated 2018-09-28 / Open, Classified / root cause: manufacturing process
TriMed Inc. began a recall of TriMed Wrist Hook Dorsal Impactor, REF IMPCT-WHD; and TriMed Wrist Hook Volar Impactor, REF IMPCT-WHV, both nonsterile, Rx, The firm name on the labels is... on . FDA classified it as Class II and gives the reason as "The WHD/WHV impactors can rust over time due to an improper manufacturing process". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0359-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Trimed Inc
- Product
- TriMed Wrist Hook Dorsal Impactor, REF IMPCT-WHD; and TriMed Wrist Hook Volar Impactor, REF IMPCT-WHV, both nonsterile, Rx, The firm name on the labels is TriMed Inc., Santa Clarita, CA.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- The WHD/WHV impactors can rust over time due to an improper manufacturing process.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The recalling firm issued letters via email on 9/28/2018 explaining the reason for recall and requesting consignees to identify potentially-affected product and quarantine it until a TriMed representative is able to conduct an inspection.
- Quantity in commerce
- 561 devices
- Distribution
- Distribution was nationwide. There was military distribution but no government distribution. Foreign distribution was made to Australia, China, South Africa, and Sweden.
- Product codes and lots
- Lot numbers 049045, 049044, and 049044R1.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0355-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0356-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0357-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0358-2020 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0359-2020 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.