Berkeley Disposa-Filter used with Berkeley VC-10 Vacuum Curettage System -
Olympus Corporation of the Americas began a recall of Berkeley Disposa-Filter used with Berkeley VC-10 Vacuum Curettage System - Product Usage: designed for rapid transcervical aspiration of the uterine cavity... on . FDA classified it as Class II and gives the reason as "Berkeley Disposa-Filter may possess an insufficient ultrasonic weld defect which joins the two-halves of the filter cartridge, thus the housing could separate". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0159-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- Berkeley Disposa-Filter used with Berkeley VC-10 Vacuum Curettage System - Product Usage: designed for rapid transcervical aspiration of the uterine cavity. Part# 54298-10/pack
- Product code
- HHI System, Abortion, Vacuum
- Reason for recall
- Berkeley Disposa-Filter may possess an insufficient ultrasonic weld defect which joins the two-halves of the filter cartridge, thus the housing could separate
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Olympus issued notification letter on 9/12/19 via Federal Express two day delivery stating reason for recall, health risk and action to take: 1. Immediately assess any affected product you have in stock, remove it from your inventory and quarantine it until it is shipped back to us. 2.Call your Olympus customer service representative at 1-888-524-7266 option 1. Olympus will issue a Return Material Authorization in order to return any affected product at no charge to you. Olympus will issue a credit or replacement to your facility for your affected product.. 4.Fax the completed Reply form to Olympus Market Quality at 484-896-7128 regardless of whether you have any affected inventory at your facility. Contact 484-896-5688 or at [email protected] for any additional information concerning this matter.
- Quantity in commerce
- 1106 boxes (10 filters/box)
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Lot number: 494604
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0159-2020 | Class II | Ongoing | Voluntary: Firm initiated |