Z-0124-2020Class II

Cerelink ICP Monitor, Model Number 826820; intended for use as an interface between compatible strain-gauge...

Integra LifeSciences Corporation / initiated 2019-09-12 / Open, Classified

Integra LifeSciences Corp. began a recall of Cerelink ICP Monitor, Model Number 826820; intended for use as an interface between compatible strain-gauge type pressure transducers and standard... on . FDA classified it as Class II and gives the reason as "There is a potential for a progressive decline in observed ICP readings of the monitor during use". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0124-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Integra LifeSciences Corporation
Product
Cerelink ICP Monitor, Model Number 826820; intended for use as an interface between compatible strain-gauge type pressure transducers and standard physiological pressure monitoring systems.
Product code
GWM Device, Monitoring, Intracranial Pressure
Reason for recall
There is a potential for a progressive decline in observed ICP readings of the monitor during use.
FDA root cause
Under Investigation by firm
Action
On 9/12/19, the customers were notified by a visit from the Sales Representative who delivered a notification letter at that time. The letter, URGENT: VOLUNTARY MEDICAL DEVICE RECALL, dated 9/12/19, included the following instructions: 1. If you have a CereLink ICP monitor, Model #826820, discontinue using the product as soon as clinically possible and remove the product from service. 2. Your sales representative will facilitate the return of your monitor. 3. Complete the information on the Acknowledgement Form attached to this letter and either provide it to your sales representative or email or fax as indicated on the form. Keep a copy of the form for your records. Should you have any questions regarding these instructions, please contact Customer Service Monday to Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873: email: [email protected] Outside USA: 781-565-1401
Quantity in commerce
189
Distribution
The products were distributed to the following US states: CA, CO, CT, FL, IN, NY, OK, PA, TX, WI, and WV. The products were distributed to the following foreign countries: Australia, Belgium.
Product codes and lots
All systems

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0124-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0124-2020Class IIOngoingVoluntary: Firm initiated