Z-0277-2020Class II

Prismaflex Control Unit, software versions below 7.21

Baxter Healthcare Corporation / initiated 2019-10-04 / Open, Classified / root cause: software

Baxter Healthcare Corporation began a recall of Prismaflex Control Unit, software versions below 7.21 on . FDA classified it as Class II and gives the reason as "Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0277-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
Prismaflex Control Unit, software versions below 7.21
Product code
KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Reason for recall
Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.
Root cause tag
software
FDA root cause
Software design
Action
Baxter Healthcare notified customers on about 10/04/2019, via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that Prismaflex Control Units with software versions 7.20 and below have the potential to produce communication errors. Instructions included that operators may continue to use the control units until the upgrade is performed, but if a communication error alarm does occur, follow the instructions presented on the graphical user interface and/or in the operators manual. It must be reinforced that it is important to manually return the extracorporeal blood to the patient, and re-train users on this process, outlined on page 10:57 of the operators manual, if necessary. A local Baxter service representative will contact your facility to determine the correction plan and schedule the software upgrade. If the affected devices was purchased from Baxter, complete the provided Baxter Customer Reply Form and return it to Baxter by faxing it to 224-270-5457 or scanning and e-mailing it to [email protected]. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. For a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures. Questions regarding this recall can be directed to Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday.
Quantity in commerce
986 units
Distribution
Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. International distribution to Australia, Bangladesh, Brunei, China, Hongkong, Indonesia, India, Japan, New Zealand, Taiwan, Korea, Singapore, Malaysia, Macau, Thailand, Vietnam, Canada, Argentina, Costa Rica, Cuba, Dominican Republic, Haiti, Barbados, Venezuela, St. Kitts, and EMEA.
Product codes and lots
Product Code: 113081; All Lot numbers; GTIN: 07332414105266

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0277-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0276-2020Class IIOngoingVoluntary: Firm initiated
Z-0277-2020Class IIOngoingVoluntary: Firm initiated
Z-0278-2020Class IIOngoingVoluntary: Firm initiated
Z-0279-2020Class IIOngoingVoluntary: Firm initiated
Z-0280-2020Class IIOngoingVoluntary: Firm initiated