Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0224-2020Class II

HYTEC Stop Solution (Allergy), 300ml bottle containing 1N sodium hydroxide solution, Part # 308051, sold...

Hycor Biomedical / initiated 2018-07-25 / Open, Classified / root cause: manufacturing process

Hycor Biomedical LLC began a recall of HYTEC Stop Solution (Allergy), 300ml bottle containing 1N sodium hydroxide solution, Part # 308051, sold separately or as a component contained in the... on . FDA classified it as Class II and gives the reason as "Some of the Stop Solution bottles, a caustic solution, have been reported to leak". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0224-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Hycor Biomedical
Product
HYTEC Stop Solution (Allergy), 300ml bottle containing 1N sodium hydroxide solution, Part # 308051, sold separately or as a component contained in the following kits: (1) HYTEC Specific/Total IgE EIA kit, Part# 74155; and (2) HYTEC Specific IgG EIA Kit, Part# 74114.
Product code
DHB System, Test, Radioallergosorbent (Rast) Immunological
Reason for recall
Some of the Stop Solution bottles, a caustic solution, have been reported to leak.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
A Field Safety Notice dated 7/25/2018 was issued via email explaining there have been reports of leakage limited to the Stop Solution bottles, but there is no impact to patient results or performance of the kits.
Quantity in commerce
185 individual bottles (170 US and 15 OUS) and 1,298/1-bottle kits (OUS)
Distribution
Stop Solution bottles were distributed to AL, CA, FL, IL, KS, LA, MS, NJ, NY, SC, TX, WA, and to the U.S. Virgin Islands. Foreign distribution of the bottles was made to Germany. There was no government/military distribution. Assay kits containing the Stop Solution bottles were distributed to FL and to the following foreign countries: Austria, Denmark, France, Germany, Greece, Italy, Kuwait, Netherlands, Russia, Spain, and United Kingdom.
Product codes and lots
Stop Solution bottles, REF 308051 - Lot numbers 154803, exp. 31-MAR-2021; and 154144, exp. 30-SEP-2020;  HYTEC Specific & Total IgE Kit, REF 74155, Lot numbers 154104-11, exp. 30-SEP-2018; and 156932-1, exp. 30-JUN-2018 - all contain Stop Solution lot 154803;  HYTEC Specific & Total IgE Kit, REF 74155, Lot numbers 154104-10, exp. 30-SEP-2018; 154104-7, exp. 30-SEP-2018; 154104-8, exp. 30-SEP-2018; and 154104-9, exp. 30 SEP-2018 - all contain Stop Solution lot 154144; and  HYTEC Specific IgG EIA ref 74114, Lot number 155324, exp. 31-DEC-2018 - contains Stop Solution lot 154144.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0224-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0224-2020Class IIOngoingVoluntary: Firm initiated