Z-0224-2020Class II
HYTEC Stop Solution (Allergy), 300ml bottle containing 1N sodium hydroxide solution, Part # 308051, sold...
Hycor Biomedical / initiated 2018-07-25 / Open, Classified / root cause: manufacturing process
Hycor Biomedical LLC began a recall of HYTEC Stop Solution (Allergy), 300ml bottle containing 1N sodium hydroxide solution, Part # 308051, sold separately or as a component contained in the... on . FDA classified it as Class II and gives the reason as "Some of the Stop Solution bottles, a caustic solution, have been reported to leak". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0224-2020
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Hycor Biomedical
- Product
- HYTEC Stop Solution (Allergy), 300ml bottle containing 1N sodium hydroxide solution, Part # 308051, sold separately or as a component contained in the following kits: (1) HYTEC Specific/Total IgE EIA kit, Part# 74155; and (2) HYTEC Specific IgG EIA Kit, Part# 74114.
- Product code
- DHB System, Test, Radioallergosorbent (Rast) Immunological
- Reason for recall
- Some of the Stop Solution bottles, a caustic solution, have been reported to leak.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- A Field Safety Notice dated 7/25/2018 was issued via email explaining there have been reports of leakage limited to the Stop Solution bottles, but there is no impact to patient results or performance of the kits.
- Quantity in commerce
- 185 individual bottles (170 US and 15 OUS) and 1,298/1-bottle kits (OUS)
- Distribution
- Stop Solution bottles were distributed to AL, CA, FL, IL, KS, LA, MS, NJ, NY, SC, TX, WA, and to the U.S. Virgin Islands. Foreign distribution of the bottles was made to Germany. There was no government/military distribution. Assay kits containing the Stop Solution bottles were distributed to FL and to the following foreign countries: Austria, Denmark, France, Germany, Greece, Italy, Kuwait, Netherlands, Russia, Spain, and United Kingdom.
- Product codes and lots
- Stop Solution bottles, REF 308051 - Lot numbers 154803, exp. 31-MAR-2021; and 154144, exp. 30-SEP-2020; HYTEC Specific & Total IgE Kit, REF 74155, Lot numbers 154104-11, exp. 30-SEP-2018; and 156932-1, exp. 30-JUN-2018 - all contain Stop Solution lot 154803; HYTEC Specific & Total IgE Kit, REF 74155, Lot numbers 154104-10, exp. 30-SEP-2018; 154104-7, exp. 30-SEP-2018; 154104-8, exp. 30-SEP-2018; and 154104-9, exp. 30 SEP-2018 - all contain Stop Solution lot 154144; and HYTEC Specific IgG EIA ref 74114, Lot number 155324, exp. 31-DEC-2018 - contains Stop Solution lot 154144.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0224-2020 | Class II | Ongoing | Voluntary: Firm initiated |