Z-0341-2020Class II

Grab n Go Opti series VIPR system Model # PRX-9632 -

Western Enterprises, Inc. / initiated 2019-08-15 / Terminated / root cause: design

Western / Scott Fetzer Company began a recall of Grab n Go Opti series VIPR system Model # PRX-9632 - Product Usage: The Grab n Go Opti system is an integrated delivery system intended to provide supplemental... on . FDA classified it as Class II and gives the reason as "Limited access to flow settings as a result of the control knob having been rotated beyond its functional range". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0341-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Western Enterprises, Inc.
Product
Grab n Go Opti series VIPR system Model # PRX-9632 - Product Usage: The Grab n Go Opti system is an integrated delivery system intended to provide supplemental oxygen, by Rx only, to patients. When administered by properly trained personnel for oxygen deficiency and resuscitation, the device is for emergency use only. For all other medical applications, the device is Rx only. The device is MR Conditional and suitable for use outside the 1,500 gauss (0.15 T) line during MR imaging for MRI systems up to 3.0 T. The device is intended for limited duration use, such as would be necessary during patient transports.
Product code
ECX Cylinder, Compressed Gas, And Valve
Reason for recall
Limited access to flow settings as a result of the control knob having been rotated beyond its functional range
Root cause tag
design
FDA root cause
Device Design
Action
Western formally communicated the product issue to Praxair on 7/12/19, who had been previously in contact with Western regarding this issue. Western plans to issue a formal communication to affected distributors in the near future. The firm has initiated corrective action in the field as of August 2019. The firm has implemented a new screening process at each fill plant to identify and isolate units that could potentially be rotated past the stop. Units which fail are isolated and returned to Western, which will inspect each failed unit, replace the failed component with the revised knob/plate, and return the refurbished unit to the fill plant. One customer specifically requested that, instead of the screening procedure above, all of its units in service be replaced with the revised design. Western fulfilled this request. On October 11, 2019, Western mailed a Medical Device Recall letter to its distributors to officially communicate the issue.
Quantity in commerce
6,790 units
Distribution
Worldwide distribution US Nationwide distributions in the states of OH, NY, PA, TX, MA, MI, KS, IL, MD, MN, WA, TN, UT, CA, NJ, LA, CT, OR, ND, RI, WI, PR, FL, ON, NV, NB, OK, IN, AZ, MS, GA, IA and countries of Costa Rica, Mexico.
Product codes and lots
All Lot manufactured from 08/31/14 to 08/07/2019

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0341-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0339-2020Class IITerminatedVoluntary: Firm initiated
Z-0340-2020Class IITerminatedVoluntary: Firm initiated
Z-0341-2020Class IITerminatedVoluntary: Firm initiated
Z-0342-2020Class IITerminatedVoluntary: Firm initiated
Z-0343-2020Class IITerminatedVoluntary: Firm initiated