Z-0168-2020Class II

Ergo 3.2mm K-wire -

Exactech, Inc. / initiated 2019-09-10 / Terminated

Exactech, Inc. began a recall of Ergo 3.2mm K-wire - Product Usage: The devices is used assist the surgeon in the implantation of Equinoxe Shoulder System components according to a... on . FDA classified it as Class II and gives the reason as "Potential for the wire to fracture at the threads and remain in the glenoid". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0168-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Exactech, Inc.
Product
Ergo 3.2mm K-wire - Product Usage: The devices is used assist the surgeon in the implantation of Equinoxe Shoulder System components according to a conventional technique for total shoulder replacement. The wire is single-use, provided sterile, and intended for transient use.
Product code
LRN Wire, Surgical
Reason for recall
Potential for the wire to fracture at the threads and remain in the glenoid.
FDA root cause
Under Investigation by firm
Action
Exactech notified customers of the recall on about 09/10/2019, via "URGENT MEDICAL DEVICE RECALL" letter. Customers were informed that Exactech had received reports of the 3.2mm K-Wire fracturing at the threads and remaining in the glenoid. Instructions included to immediately cease distribution of or use of the product, provide the recall information to accounts that may have the affected product in their possession, identify and quarantine any affected products in inventory, and complete and return the attached Recall Acknowledgement Form and Recall Inventory Response form to Exactech via email at [email protected].
Quantity in commerce
158 devices
Distribution
US Nationwide including the states of AR, AZ, CA, CO, FL, IL, IN, KS, LA, MN, NY, NY, OH, SC, TN, TX, VA, WA.
Product codes and lots
Catalog Number: 321-52-06, All Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0168-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0168-2020Class IITerminatedVoluntary: Firm initiated