SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system
Siemens Medical Solutions USA, Inc began a recall of SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system on . FDA classified it as Class II and gives the reason as "SOMATOM go.Top (Models #11061640 & 11061648) with software syngo CT VA20A_SP2 and active Guide&GO license No dose documentation and no Dose Alert for the special mode i- Sequence during interventional procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0265-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- SOMATOM go.Top (Models #11061640 & 11061648) with software syngo CT VA20A_SP2 and active Guide&GO license No dose documentation and no Dose Alert for the special mode i- Sequence during interventional procedures
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens issued a Customer Safety Advisory Notice to customers via CT072/19 on 9/24/2019. The letter identifies the problem, health risk and action to take: To avoid this issue, please refrain from using the i-Sequence scan mode in combination with the 3x3 mm collimation until this issue has been solved. Please use the i-Sequence with the 3x5 mm collimation instead, as here the correct dose information will be shown. This issue will be resolved with an improved software version syngo CT VA20A_SP2b. The software update will be released, free of charge, with the update CT073/19/S for all affected systems. The software update release is planned for Q3/2019. All future service packs will contain the update as well. Questions: contact service organization at 1-800-888-7436.
- Quantity in commerce
- 14 US
- Distribution
- US in the states of: GA, IN, MO, MS, NJ, OH, PA
- Product codes and lots
- software syngo CT VA20A_SP2 and active Guide&GO license
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0265-2020 | Class II | Terminated | Voluntary: Firm initiated |