Z-0265-2020Class II

SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system

Siemens Medical Solutions USA, Inc. / initiated 2019-09-24 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system on . FDA classified it as Class II and gives the reason as "SOMATOM go.Top (Models #11061640 & 11061648) with software syngo CT VA20A_SP2 and active Guide&GO license No dose documentation and no Dose Alert for the special mode i- Sequence during interventional procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0265-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
SOMATOM go.Top, Models 11061640 & 11061648- a Computer tomography x-ray system
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
SOMATOM go.Top (Models #11061640 & 11061648) with software syngo CT VA20A_SP2 and active Guide&GO license No dose documentation and no Dose Alert for the special mode i- Sequence during interventional procedures
Root cause tag
software
FDA root cause
Software design
Action
Siemens issued a Customer Safety Advisory Notice to customers via CT072/19 on 9/24/2019. The letter identifies the problem, health risk and action to take: To avoid this issue, please refrain from using the i-Sequence scan mode in combination with the 3x3 mm collimation until this issue has been solved. Please use the i-Sequence with the 3x5 mm collimation instead, as here the correct dose information will be shown. This issue will be resolved with an improved software version syngo CT VA20A_SP2b. The software update will be released, free of charge, with the update CT073/19/S for all affected systems. The software update release is planned for Q3/2019. All future service packs will contain the update as well. Questions: contact service organization at 1-800-888-7436.
Quantity in commerce
14 US
Distribution
US in the states of: GA, IN, MO, MS, NJ, OH, PA
Product codes and lots
software syngo CT VA20A_SP2 and active Guide&GO license

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0265-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0265-2020Class IITerminatedVoluntary: Firm initiated