Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0219-2020Class II

Model No. 01-1503-0032 (Adolescent Nail Attachment Bolt). Component of the Pediatric Nailing Platform

Orthopediatrics Corp / initiated 2019-08-19 / Terminated / root cause: manufacturing process

OrthoPediatrics Corp began a recall of Model No. 01-1503-0032 (Adolescent Nail Attachment Bolt). Component of the Pediatric Nailing Platform. Product Usage: an orthopedic implant system commonly... on . FDA classified it as Class II and gives the reason as "Binding can occur between the attachment bolt and intramedullary nail, which may lead to delays in surgical procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0219-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Orthopediatrics Corp
Product
Model No. 01-1503-0032 (Adolescent Nail Attachment Bolt). Component of the Pediatric Nailing Platform. Product Usage: an orthopedic implant system commonly referred to as an intramedullary fixation rod (product code HSB). The attachment bolts are one component of the surgical instruments within the system.
Product code
HSB Rod, Fixation, Intramedullary And Accessories
Reason for recall
Binding can occur between the attachment bolt and intramedullary nail, which may lead to delays in surgical procedures.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On August 19, 2019, the firm notified customers of the issue via emailed notice titled Urgent Medical Device Recall. Customers were informed of the issue of binding between the attachment bolt and the intramedullary nail. The firm asked customers to return all attachment bolts in the field. The issue does not impact intramedullary nails that have been implanted. No additional corrective actions need to be taken for product already implanted. Any questions regarding this recall can be directed to Kriss Anderson, VP of Quality Systems and Regulatory Affairs, at OrthoPediatrics at [email protected] or 574-267-0897.
Quantity in commerce
969 total
Distribution
Worldwide distributions - US Nationwide distributions including states of AL, AZ, CA, CO, FL, GA, IA, IL, IN, KY, MD, MI, NC, NY, OH, OK, TN, TX, UT, VA, WA, and WI. Country of Australia.
Product codes and lots
175221-A 177577-A 183280-A 187375-A 189808-A 192279-B 192432-B 198082-C

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0219-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0218-2020Class IITerminatedVoluntary: Firm initiated
Z-0219-2020Class IITerminatedVoluntary: Firm initiated