Model No. 01-1503-0032 (Adolescent Nail Attachment Bolt). Component of the Pediatric Nailing Platform
OrthoPediatrics Corp began a recall of Model No. 01-1503-0032 (Adolescent Nail Attachment Bolt). Component of the Pediatric Nailing Platform. Product Usage: an orthopedic implant system commonly... on . FDA classified it as Class II and gives the reason as "Binding can occur between the attachment bolt and intramedullary nail, which may lead to delays in surgical procedures". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0219-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Orthopediatrics Corp
- Product
- Model No. 01-1503-0032 (Adolescent Nail Attachment Bolt). Component of the Pediatric Nailing Platform. Product Usage: an orthopedic implant system commonly referred to as an intramedullary fixation rod (product code HSB). The attachment bolts are one component of the surgical instruments within the system.
- Product code
- HSB Rod, Fixation, Intramedullary And Accessories
- Reason for recall
- Binding can occur between the attachment bolt and intramedullary nail, which may lead to delays in surgical procedures.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On August 19, 2019, the firm notified customers of the issue via emailed notice titled Urgent Medical Device Recall. Customers were informed of the issue of binding between the attachment bolt and the intramedullary nail. The firm asked customers to return all attachment bolts in the field. The issue does not impact intramedullary nails that have been implanted. No additional corrective actions need to be taken for product already implanted. Any questions regarding this recall can be directed to Kriss Anderson, VP of Quality Systems and Regulatory Affairs, at OrthoPediatrics at [email protected] or 574-267-0897.
- Quantity in commerce
- 969 total
- Distribution
- Worldwide distributions - US Nationwide distributions including states of AL, AZ, CA, CO, FL, GA, IA, IL, IN, KY, MD, MI, NC, NY, OH, OK, TN, TX, UT, VA, WA, and WI. Country of Australia.
- Product codes and lots
- 175221-A 177577-A 183280-A 187375-A 189808-A 192279-B 192432-B 198082-C
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0218-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0219-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY