Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0505-2020Class II

XCELA, 8F SINGLE PLASTIC PORT FILLED, VALVED, UPN H965440400, Cat. No. 44-040 -

Boston Scientific Corporation / initiated 2019-09-18 / Terminated

Angiodynamics Inc. (Navilyst Medical Inc.) began a recall of XCELA, 8F SINGLE PLASTIC PORT FILLED, VALVED, UPN H965440400, Cat. No. 44-040 - Product Usage: Long-term access to the central venous system for administration... on . FDA classified it as Class II and gives the reason as "Snap lock connectors provided within implantable port kits may not meet dimensional specifications, and the catheter cannot be inserted into the snap lock connector prior to connection to the port chamber, resulting in...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0505-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Boston Scientific Corporation
Product
XCELA, 8F SINGLE PLASTIC PORT FILLED, VALVED, UPN H965440400, Cat. No. 44-040 - Product Usage: Long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, chemotherapy, analgesics, nutritional therapy and blood products. The device is also indicated for blood specimen withdrawal.
Product code
LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
Reason for recall
Snap lock connectors provided within implantable port kits may not meet dimensional specifications, and the catheter cannot be inserted into the snap lock connector prior to connection to the port chamber, resulting in a delay to the implantation procedure.
FDA root cause
Under Investigation by firm
Action
On September 18, 2019, AngioDynamics sent Urgent Voluntary Medical Device Letters to impacted consignees. Customers were asked to do the following: IMMEDIATELY o Stop using the product subject to recall. o Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). o Segregate this product in a secure location for return to AngioDynamics, Inc. o Forward a copy of this recall notification to all sites to which you have distributed affected product. IN THE EVENT THAT THE PORT PRODUCT UNDER RECALL HAS BEEN UTILIZED (IMPLANTED) o The occurrence rate of an out of specification snap lock connector is less than 100%. o A port kit containing a properly sized snap lock connector allows the catheter to be securely attached to the port. o There are no actions required for an implanted port with a properly attached catheter. Customers were asked to complete and return the Reply Verification Form, and to package and return the recalled product.
Quantity in commerce
582
Distribution
Worldwide distribution - US Nationwide in the state of CA, CT, FL, GA, IA, IN, KS, KY, MI, MO, MS, MT, NE, NJ, NY, OH, OK, OR, PA, TN, TX, WA, WI and country of Canada.
Product codes and lots
5480564

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0505-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0501-2020Class IITerminatedVoluntary: Firm initiated
Z-0502-2020Class IITerminatedVoluntary: Firm initiated
Z-0503-2020Class IITerminatedVoluntary: Firm initiated
Z-0504-2020Class IITerminatedVoluntary: Firm initiated
Z-0505-2020Class IITerminatedVoluntary: Firm initiated