XCELA, 8F SINGLE PLASTIC PORT FILLED, VALVED, UPN H965440400, Cat. No. 44-040 -
Angiodynamics Inc. (Navilyst Medical Inc.) began a recall of XCELA, 8F SINGLE PLASTIC PORT FILLED, VALVED, UPN H965440400, Cat. No. 44-040 - Product Usage: Long-term access to the central venous system for administration... on . FDA classified it as Class II and gives the reason as "Snap lock connectors provided within implantable port kits may not meet dimensional specifications, and the catheter cannot be inserted into the snap lock connector prior to connection to the port chamber, resulting in...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0505-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Boston Scientific Corporation
- Product
- XCELA, 8F SINGLE PLASTIC PORT FILLED, VALVED, UPN H965440400, Cat. No. 44-040 - Product Usage: Long-term access to the central venous system for administration of fluids including but not limited to hydration fluids, chemotherapy, analgesics, nutritional therapy and blood products. The device is also indicated for blood specimen withdrawal.
- Product code
- LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
- Reason for recall
- Snap lock connectors provided within implantable port kits may not meet dimensional specifications, and the catheter cannot be inserted into the snap lock connector prior to connection to the port chamber, resulting in a delay to the implantation procedure.
- FDA root cause
- Under Investigation by firm
- Action
- On September 18, 2019, AngioDynamics sent Urgent Voluntary Medical Device Letters to impacted consignees. Customers were asked to do the following: IMMEDIATELY o Stop using the product subject to recall. o Remove any affected (recalled) product from your inventory (whether in labs, Central Supply, Shipping and Receiving or ANY other location). o Segregate this product in a secure location for return to AngioDynamics, Inc. o Forward a copy of this recall notification to all sites to which you have distributed affected product. IN THE EVENT THAT THE PORT PRODUCT UNDER RECALL HAS BEEN UTILIZED (IMPLANTED) o The occurrence rate of an out of specification snap lock connector is less than 100%. o A port kit containing a properly sized snap lock connector allows the catheter to be securely attached to the port. o There are no actions required for an implanted port with a properly attached catheter. Customers were asked to complete and return the Reply Verification Form, and to package and return the recalled product.
- Quantity in commerce
- 582
- Distribution
- Worldwide distribution - US Nationwide in the state of CA, CT, FL, GA, IA, IN, KS, KY, MI, MO, MS, MT, NE, NJ, NY, OH, OK, OR, PA, TN, TX, WA, WI and country of Canada.
- Product codes and lots
- 5480564
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0501-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0502-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0503-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0504-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0505-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.