Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0674-2020Class II

Philips Volcano FFR software used in the following systems connected to the McKesson, Schwarzer, and some...

Volcano Corporation / initiated 2018-11-07 / Terminated / root cause: manufacturing process

Volcano Corporation began a recall of Philips Volcano FFR software used in the following systems connected to the McKesson, Schwarzer, and some Philips Xper Hemodynamic Systems (HMS): (1) Volcano... on . FDA classified it as Class II and gives the reason as "Interoperability issue that affects certain systems that use the recalled software when it is connected to specific Hemodynamic Systems (HMS)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0674-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Volcano Corporation
Product
Philips Volcano FFR software used in the following systems connected to the McKesson, Schwarzer, and some Philips Xper Hemodynamic Systems (HMS): (1) Volcano Imaging System s5i, Part #807400001; (2) s5ix Imaging System with s5iz PC Workstation Refurbished, Part#807400001 R and 807401200; (3) s5i Imaging System, Part#806300003; (4) s5i Imaging System, Part#5322466; (5) s5x Imaging System, Part#807300001; (6) s5x Imaging System Refurbished, Pat#807300001 R; (7) s5 Imaging System Refurbished, Part#804200001 R; (8) CORE Mobile Imaging System (120V), Part #400-0100.01; (9) CORE Mobile Imaging System Refurbished, Part #400-0100.01-R; (10) CORE Mobile Imaging System (240V), Part #400-0100.07; (11) CORE Mobile Imaging System Refurbished, Part #400-0100.07-R; (12) CORE Mobile Imaging System (100V), Part #400-0100.08; (13) CORE Mobile Imaging System Refurbished, Part #400-0100.08-R; and (14) CORE Imaging System, Part #400-0100.02; (15) CORE Integrated Imaging System Refurbished, Part#400 0100.02 R.
Product code
IYO System, Imaging, Pulsed Echo, Ultrasonic
Reason for recall
Interoperability issue that affects certain systems that use the recalled software when it is connected to specific Hemodynamic Systems (HMS).
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
The recalling firm issued letters dated 10/26/2018 during the week of 11/7/2018. The letters explain how to detect any potential interoperability and provide details of the impact to the system performance. The letters will be sent via proof of delivery. Field Service Engineers will visit each site to install the software update. Urgent Medical Device Field Action notices, dated 04/27/2020, were mailed to customers during the week of 04/27/2020. Customers informed of the following: If the calibrated device Pa values have a misalignment greater than plus/minus 3mm Hg, contact Technical Support so that the system can be corrected. Additionally, a software update has been developed which will resolve the interoperability issue. The Technical support team will deploy the software update on each impacted system to address the interoperability issue described above. Please ensure that a copy of this letter is provided to all personnel within your organization who handle this device. Customers with additional questions are encouraged to call: 18002284728, Option 2, or email: [email protected]
Quantity in commerce
3,667 units with software
Distribution
Distribution was nationwide. There was government/military distribution. Foreign distribution was made to Canada, Austria, Australia, Badajoz, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Ireland, Israel, Italy, Jordan, Japan, Korea, Republic of, Lebanon, Lithuania, Malaysia, Malta, Mexico, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Sierra Leone, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
Product codes and lots
FFR V2.5 Software Kit, Catalog 4536-010-57091 (Legacy 435-0602.48) and FFR V2.5 Update Installation Disk, Catalog 4536-010-52781 (Legacy 435-0100.98)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0674-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0674-2020Class IITerminatedVoluntary: Firm initiated