Platelet Aggregation Kit, Cat. No.5369, For use in platelet aggregation studies, Contains: 2X1 mL ADP Reagent...
Helena Laboratories, Corp. began a recall of Platelet Aggregation Kit, Cat. No.5369, For use in platelet aggregation studies, Contains: 2X1 mL ADP Reagent containing adenosine diphosphate, 2X 1 mL... on . FDA classified it as Class III and gives the reason as "Through an investigation, it has been determined that vials of Collagen packaged within the Collagen Reagent and Platelet Aggregation Kit may contain particulate matter or microbial growth". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0536-2020
- Classification
- Class III
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Helena Laboratories, Corp.
- Product
- Platelet Aggregation Kit, Cat. No.5369, For use in platelet aggregation studies, Contains: 2X1 mL ADP Reagent containing adenosine diphosphate, 2X 1 mL Collagen Reagent containing collagen (equine tendon), 2 x 1 mL Epinephrine Reagent containing L-epinephrine, FOR IN-VITRO DIAGNOSTIC USE, Store at 2 to 8 C, 0011724/14(3)
- Product code
- GHR Reagent, Platelet Aggregation
- Reason for recall
- Through an investigation, it has been determined that vials of Collagen packaged within the Collagen Reagent and Platelet Aggregation Kit may contain particulate matter or microbial growth.
- Root cause tag
- packaging
- FDA root cause
- Unknown/Undetermined by firm
- Action
- On 05/23/2018, the firm's technical services contacted consignees via phone call with a email follow-up of their recall notification letter. Consignees were advised of the following: 1. Prevent any shipment or schedule demonstrations with these lots 2. Remove the lots from available inventory and discard COLLAGEN vials. 3. Verify amount REMOVED FROM STOCK 4. COMPLETE THE FORM AND RETURN TO HELENA LABORATORIES, USA VIA: fax: 409-842-1874 OR E-MAIL TO [email protected]. For questions, contact Jessica W. Jones, MD, MPH - Mgr., Technical Services at: 1 (409) 842-3714 ext. 1177
- Quantity in commerce
- 383 Collagen Kits
- Distribution
- US:TX, WV, OH, TN, GA, TN, NY, ME, MA,LA, AZ OUS: Israel, Korea, Australia, Indonesia, Canada, Thailand, Colombia, Brasil , China , Taiwan
- Product codes and lots
- Component Lot # 5-17-5368 (Collagen), Exp. 5/31/19 Kit Lot # 3-17-5369 (Aggregation kit containing Collagen), Exp. 11/30/18 Kit Lot # 5-17-5368 (Aggregation Kit containing Collagen), Exp. 05/31/2019
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0536-2020 | Class III | Ongoing | Voluntary: Firm initiated | ||
| Z-0537-2020 | Class III | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
