Cochlear Baha 5 Sound Processor Tamper Proof Battery Door (3 pcs.)

Cochlear Americas / initiated 2018-12-18 / Terminated / root cause: packaging

Cochlear Americas Inc. began a recall of Cochlear Baha 5 Sound Processor Tamper Proof Battery Door (3 pcs.) on . FDA classified it as Class II and gives the reason as "Product was packaged with the Incorrect tamper proof battery door". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0792-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cochlear Americas
Product
Cochlear Baha 5 Sound Processor Tamper Proof Battery Door (3 pcs.)
Product code
LXB Hearing Aid, Bone Conduction
Reason for recall
Product was packaged with the Incorrect tamper proof battery door.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
On 12/18/2018, the firm sent a "MEDICAL DEVICE RECALL NOTICE" to affected customers via telephone or email informing them of a packaging error for a specific lot (LOT # COH1253733) Tamper Proof Battery Doors. Customers are instructed to: 1-Examine their inventory and if they should have any of the affected product in their possession to carefully verify that they are the tamper proof version, and if in doubt about the functionality or identity - to scrap the item and to inform the Recalling Firm by filling out and return the Response Form. The Recalling Firm will arrange to send replacement product. 2-If customers receive complaints or queries from their users about "loose battery doors" or similar, to please advise the user to check the functionality according to the Recalling Fir's instructions "Baha 5 User Manual" and to scrap and replace as necessary. 3-To identify customers or users to whom they have distributed battery doors from the affected lot and to provide contact information to the Recalling Firm. The Recalling Firm will make contact, explain the issue, and send replacement product directly to them if necessary. NOTE: If you (the customer) do not have specific records showing with customers may have received affected product, they are to consider providing the Recalling Firm with a list of all customers that received a Baha 5 system between Sept. 1, 2018 and Dec. 18, 2018 so that the Recalling Firm may contact them. For any questions about the product recall customers can contact the Recalling Firm's Quality Department at 303-264-2163.
Quantity in commerce
50 units
Distribution
Domestic: CA, FL, IL, IN, MA, MO, NY, SD, TN, and TX OUS Regions; Asia Pacific (APAC), and Europe, The Middle East and Africa (EMEA)
Product codes and lots
Part Number: P1129017 UDI Number: 09321502041096 Lot Number: COH1253733

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0792-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0792-2020Class IITerminatedVoluntary: Firm initiated