Z-0250-2020Class II

Cordis MAXI LD PTA Dilatation Catheter, for cardiovascular use

Cordis US Corp. / initiated 2019-10-04 / Terminated / root cause: manufacturing process

Cordis Corporation began a recall of Cordis MAXI LD PTA Dilatation Catheter, for cardiovascular use on . FDA classified it as Class II and gives the reason as "Product was manufactured utilizing an expired inner body". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0250-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cordis US Corp.
Product
Cordis MAXI LD PTA Dilatation Catheter, for cardiovascular use.
Product code
LIT Catheter, Angioplasty, Peripheral, Transluminal
Reason for recall
Product was manufactured utilizing an expired inner body.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Cordis notified customers on about 10/04/2019, via Urgent MEDICAL DEVICE RECALL letter delivered by sales representatives. Customers were informed that the Cordis MAXI LD" PTA Dilatation Catheter was manufactured utilizing an expired inner body. For an expired inner body, a degradation of material may contribute to a collapse of the guidewire lumen resulting in a resistance during insertion or backloading of the guidewire into the lumen. Instructions included to read the recall letter, immediately check inventory to conform if the affected product is in your possession, identify and set aside any units from the affected lot, complete and return the provided Acknowledgement Form directly to Cordis via fax (614-495-5571) or email to [email protected]. Customers were also instructed to arrange for return of affected product, provide the recall letter to personnel at your facility who need to be made aware and any other facility that may have been sent the affected units.
Quantity in commerce
21 units
Distribution
US distribution to AL, AZ, FL, MN, and MO
Product codes and lots
Catalog Number: 4171460S; Lot Number: 82169860; Expiry Date: 10/31/2020

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0250-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0250-2020Class IITerminatedVoluntary: Firm initiated