Z-0250-2020Class II
Cordis MAXI LD PTA Dilatation Catheter, for cardiovascular use
Cordis US Corp. / initiated 2019-10-04 / Terminated / root cause: manufacturing process
Cordis Corporation began a recall of Cordis MAXI LD PTA Dilatation Catheter, for cardiovascular use on . FDA classified it as Class II and gives the reason as "Product was manufactured utilizing an expired inner body". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0250-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cordis US Corp.
- Product
- Cordis MAXI LD PTA Dilatation Catheter, for cardiovascular use.
- Product code
- LIT Catheter, Angioplasty, Peripheral, Transluminal
- Reason for recall
- Product was manufactured utilizing an expired inner body.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Cordis notified customers on about 10/04/2019, via Urgent MEDICAL DEVICE RECALL letter delivered by sales representatives. Customers were informed that the Cordis MAXI LD" PTA Dilatation Catheter was manufactured utilizing an expired inner body. For an expired inner body, a degradation of material may contribute to a collapse of the guidewire lumen resulting in a resistance during insertion or backloading of the guidewire into the lumen. Instructions included to read the recall letter, immediately check inventory to conform if the affected product is in your possession, identify and set aside any units from the affected lot, complete and return the provided Acknowledgement Form directly to Cordis via fax (614-495-5571) or email to [email protected]. Customers were also instructed to arrange for return of affected product, provide the recall letter to personnel at your facility who need to be made aware and any other facility that may have been sent the affected units.
- Quantity in commerce
- 21 units
- Distribution
- US distribution to AL, AZ, FL, MN, and MO
- Product codes and lots
- Catalog Number: 4171460S; Lot Number: 82169860; Expiry Date: 10/31/2020
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0250-2020 | Class II | Terminated | Voluntary: Firm initiated |