Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of...
Philips Medical Systems (Cleveland) Inc began a recall of Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the... on . FDA classified it as Class I and gives the reason as "An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0214-2020
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.
- Product code
- KPS System, Tomography, Computed, Emission
- Reason for recall
- An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On September 19, 2019 the firm distributed Urgent Field Safety Notice (FSN) Medical Device Correction letters by direct communication from Philips Field Service Engineers instructing them to: - Discontinue system use until further notice - Inform those who need to be aware within your organization or any organization where the potentially affected devices have been transferred (If appropriate). - Maintain this notice with your system Instructions for Use (IFU) until the correction is made to the system. ACTIONS PLANNED BY PHILIPS: Philips Healthcare is distributing this FSN to all affected customers/users and will deploy a solution addressing the issue upon completion of the investigation. FURTHER INFORMATION AND SUPPORT: If you need any further information or support concerning this issue, please contact your local Philips representative: For SUPPORT in North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, follow the prompts).
- Quantity in commerce
- 1167 in total
- Distribution
- Nationwide Distribution.
- Product codes and lots
- System Code: 882290 and 882291 All systems.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0213-2020 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0214-2020 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0215-2020 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0216-2020 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0217-2020 | Class I | Terminated | Voluntary: Firm initiated |