Z-0216-2020Class I

Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of...

Philips North America LLC / initiated 2019-09-19 / Terminated / root cause: manufacturing process

Philips Medical Systems (Cleveland) Inc began a recall of Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the... on . FDA classified it as Class I and gives the reason as "An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0216-2020
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Forte Gamma Camera System is intended to produce images depicting the anatomical distribution of single-photon and positron-emitting radioisotopes within the body for interpretation by medical personnel.
Product code
KPS System, Tomography, Computed, Emission
Reason for recall
An issue with the Detector for the Forte Family of cameras may result in either detector 1 or detector 2 falling unimpeded vertically to the end stops of its travel limit.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On September 19, 2019 the firm distributed Urgent Field Safety Notice (FSN) Medical Device Correction letters by direct communication from Philips Field Service Engineers instructing them to: - Discontinue system use until further notice - Inform those who need to be aware within your organization or any organization where the potentially affected devices have been transferred (If appropriate). - Maintain this notice with your system Instructions for Use (IFU) until the correction is made to the system. ACTIONS PLANNED BY PHILIPS: Philips Healthcare is distributing this FSN to all affected customers/users and will deploy a solution addressing the issue upon completion of the investigation. FURTHER INFORMATION AND SUPPORT: If you need any further information or support concerning this issue, please contact your local Philips representative: For SUPPORT in North America and Canada, contact the Customer Care Solutions Center (1-800-722-9377, follow the prompts).
Quantity in commerce
1167 in total
Distribution
Nationwide Distribution.
Product codes and lots
System Code: 889456 System Serial Number: 24

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0216-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0213-2020Class ITerminatedVoluntary: Firm initiated
Z-0214-2020Class ITerminatedVoluntary: Firm initiated
Z-0215-2020Class ITerminatedVoluntary: Firm initiated
Z-0216-2020Class ITerminatedVoluntary: Firm initiated
Z-0217-2020Class ITerminatedVoluntary: Firm initiated