lmpella CP with Smart Assist, Model Number: 0048-0003. Left heart support blood pump, for Use During...
Abiomed, Inc. began a recall of lmpella CP with Smart Assist, Model Number: 0048-0003. Left heart support blood pump, for Use During Cardiogenic Shock and High-Risk PCI on . FDA classified it as Class II and gives the reason as "This one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0345-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abiomed, Inc.
- Product
- lmpella CP with Smart Assist, Model Number: 0048-0003. Left heart support blood pump, for Use During Cardiogenic Shock and High-Risk PCI
- Product code
- OZD Temporary Non-Roller Type Left Heart Support Blood Pump
- Reason for recall
- This one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period. The pump may have migrated out of position and possibly caused injury. The positioning and suction alarms were also disabled.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- The affected device was retrieved from the only customer on 9/4/19. An Urgent Medical Device Recall notification letter dated 9/12/19 was sent to that customer as well.
- Quantity in commerce
- 1
- Distribution
- The product was distributed to GA
- Product codes and lots
- UDI: (01)00813502011371(10)1430951(17)210331(21)192148A; Serial Number: 192148A
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0345-2020 | Class II | Terminated | Voluntary: Firm initiated |