Cyberonics VNS Therapy AspireHC Model 105 Generator, 26-0007-4000/3. The firm name on the label is...
LivaNova USA Inc began a recall of Cyberonics VNS Therapy AspireHC Model 105 Generator, 26-0007-4000/3. The firm name on the label is CYBERONICS, INC., Houston, TX. Vagus nerve stimulator on . FDA classified it as Class II and gives the reason as "This recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0868-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- Cyberonics VNS Therapy AspireHC Model 105 Generator, 26-0007-4000/3. The firm name on the label is CYBERONICS, INC., Houston, TX. Vagus nerve stimulator.
- Product code
- LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy
- Reason for recall
- This recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators. *Note this recall occurred in November 2018.
- Root cause tag
- manufacturing process
- FDA root cause
- Manufacturing material removal
- Action
- The recalling firm issued Dear Doctor letters dated 11/16/2018 explaining the reason for the voluntary correction and pointing out a correction was previously initiated in June, 2017, related to this issue for devices distributed prior to October, 2015.
- Quantity in commerce
- 106 devices
- Distribution
- Distribution US nationwide, including Puerto Rico. There was no foreign/military/government consignees.
- Product codes and lots
- Serial numbers 35794, 37762, 37792, 37872, 37874, 37881, 37885, 37889, 37890, 37893, 37913, 37915, 37916, 37965, 37975, 37977, 37984, 37987, 37988, 37990, 38051, 38054, 38068, 38132, 38148, 38153, 38154, 38155, 38162, 38173, 38210, 38224, 38226, 38239, 38240, 38243, 38245, 38246, 38248, 38249, 38252, 38268, 38272, 38273, 38274, 38275, 38277, 38279, 38282, 38284, 38292, 38293, 38296, 38299, 38302, 38306, 38308, 38312, 38314, 38315, 38319, 38321, 38326, 38330, 38344, 38443, 38464, 38490, 38505, 38517, 38529, 38530, 38549, 38561, 38564, 38565, 38625, 38628, 38651, 38661, 38689, 38691, 38698, 38702, 38719, 38722, 38730, 38735, 38745, 38751, 38754, 38755, 38771, 38773, 38775, 38779, 38800, 38804, 38805, 38809, 39334, 39344, 39347, 39354, 39362, and 39731. The expiration date range is 2/25/2017-5/1/2017.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0868-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0869-2020 | Class II | Terminated | Voluntary: Firm initiated |