Z-0755-2020Class I

CARESCAPE Respiratory Module E-sCOVX

GE Healthcare / initiated 2019-09-27 / Terminated / root cause: manufacturing process

GE Healthcare, LLC began a recall of CARESCAPE Respiratory Module E-sCOVX on . FDA classified it as Class I and gives the reason as "Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0755-2020
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
CARESCAPE Respiratory Module E-sCOVX
Product code
CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase
Reason for recall
Displayed End Tidal Oxygen (EtO2) and Fraction of Inspired Oxygen (FiO2) values may be incorrect which could cause up to 50% measurement error in the EtO2/FiO2 values.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm issued the Urgent Medical Device Correction Letter dated 09/27/2019. The letter describes the failure and requests replacement of the affected module with another module. If a replacement is not available, the consignee is directed to follow the instructions below to ensure that your respiratory module O2 reading is within specifications (a) before any new patient case is started and (b) for prolonged patient cases, at a minimum of once a day: 1) Connect the module to the host device; 2) Ensure the module is fully warmed up by having it powered on for minimum 20 minutes; 3) Configure the O2 reading and waveform to the host device screen as defined in the host device instructions; 4) Feed 100% O2 gas flow from the ventilator to the module and observe the O2 reading and waveform; and 5) Ensure the O2 reading is within specifications 100% +/-3% and there is no erratic behavior in the waveform. 6) If you are unable to complete the above steps successfully, remove the respiratory module from clinical use and quarantine. Contact GE Healthcare Service or your local Service Representative. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Quantity in commerce
3 units
Distribution
US OUS: Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Korea (Republic of), Malaysia, Mexico, Myanmar, Netherlands, New Zealand, Peru, Poland, Republic of Serbia, Romania, Russia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Uganda, and Unites Kingdom
Product codes and lots
Mfg. Lot or Serial # UDI # SGV19276034HA 010084068210428921SGV19276034HA11190702 SGV19266019HA 010084068210428921SGV19266019HA11190626 SGV14404046HA Not Applicable

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0755-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0754-2020Class ITerminatedVoluntary: Firm initiated
Z-0755-2020Class ITerminatedVoluntary: Firm initiated
Z-0756-2020Class ITerminatedVoluntary: Firm initiated
Z-0757-2020Class ITerminatedVoluntary: Firm initiated
Z-0758-2020Class ITerminatedVoluntary: Firm initiated
Z-0759-2020Class ITerminatedVoluntary: Firm initiated
Z-0760-2020Class ITerminatedVoluntary: Firm initiated