Z-0412-2020Class II

Randox Liquid Cardiac Controls Catalogue Number CQ5051

Randox Laboratories, Limited / initiated 2019-10-04 / Terminated

Randox Laboratories, Limited began a recall of Randox Liquid Cardiac Controls Catalogue Number CQ5051 on . FDA classified it as Class II and gives the reason as "Removal of claims for N-proBNP in the Liquid Cardiac Control model numbers CQ5051 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0412-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Randox Laboratories, Limited
Product
Randox Liquid Cardiac Controls Catalogue Number CQ5051
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
Removal of claims for N-proBNP in the Liquid Cardiac Control model numbers CQ5051 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots.
FDA root cause
Under Investigation by firm
Action
On October 1, 2019 the firm distributed Urgent Medical Device Recall letters by e-mail instructing them to: - Inspect your stock and quarantine affected stock. - Replace the value sheet in the kit with the revised value sheet provided. - Randox is not recommending a review of previous results as changes in quality control recovery would be reviewed at the time of occurrence. - Discuss the contents of this notice with your Medical Director. - Complete and return the response form 12187-QA to [email protected] within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization.
Quantity in commerce
6,208
Distribution
US Distribution: WV, CA, DE, NJ, NY, GA, IN, PR OUS Distribution: CA, PL, CH, JP, NL, IT, CN, GB, IE, RU, MD, BY, SA, VN, AE, BH, DK, CO, ES, FR, TW, PH, TH, OM, HU, IN, KR, ZA, AU, CZ, PE, CL, IR, NZ, BE, DE, BA, SV, SK, HK, PT, PK, BR, TR, UA, SE, QA, LT, PE,
Product codes and lots
Catalogue Number: CQ5051 GTIN: 05055273207446 Batch/lot: 4243CK (Exp. 28Nov2019), 4246CK (Exp. 28Nov2019), 4249CK (Exp. 28Nov2019), 4260CK (Exp. 28Nov2019), 4311CK (Exp. 28May2020), 4317CK (Exp. 28May2020)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0412-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0412-2020Class IITerminatedVoluntary: Firm initiated
Z-0413-2020Class IITerminatedVoluntary: Firm initiated
Z-0414-2020Class IITerminatedVoluntary: Firm initiated