Z-0413-2020Class II
Randox Liquid Cardiac Controls Catalogue Number CQ5052
Randox Laboratories, Limited / initiated 2019-10-04 / Terminated
Randox Laboratories, Limited began a recall of Randox Liquid Cardiac Controls Catalogue Number CQ5052 on . FDA classified it as Class II and gives the reason as "Removal of claims for N-proBNP in the Liquid Cardiac Control catalogue number CQ5052 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0413-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Randox Laboratories, Limited
- Product
- Randox Liquid Cardiac Controls Catalogue Number CQ5052.
- Product code
- JJY Multi-Analyte Controls, All Kinds (Assayed)
- Reason for recall
- Removal of claims for N-proBNP in the Liquid Cardiac Control catalogue number CQ5052 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots.
- FDA root cause
- Under Investigation by firm
- Action
- On October 1, 2019 the firm distributed Urgent Medical Device Recall letters by e-mail instructing them to: - Inspect your stock and quarantine affected stock. - Replace the value sheet in the kit with the revised value sheet provided. - Randox is not recommending a review of previous results as changes in quality control recovery would be reviewed at the time of occurrence. - Discuss the contents of this notice with your Medical Director. - Complete and return the response form 12187-QA to [email protected] within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization.
- Quantity in commerce
- 7,505
- Distribution
- US Distribution: WV, CA, DE, NJ, NY, GA, IN, PR OUS Distribution: CA, PL, CH, JP, NL, IT, CN, GB, IE, RU, MD, BY, SA, VN, AE, BH, DK, CO, ES, FR, TW, PH, TH, OM, HU, IN, KR, ZA, AU, CZ, PE, CL, IR, NZ, BE, DE, BA, SV, SK, HK, PT, PK, BR, TR, UA, SE, QA, LT, PE,
- Product codes and lots
- Catalogue Number: CQ5052 GTIN: 05055273207453 Batch/lot: 4244CK (Exp. 28Nov2019), 4247CK (Exp. 28Nov2019), 4261CK (Exp. 28Nov2019), 4312CK Exp. 28Jun2020) and 4315CK (Exp. 28Jun2020)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0412-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0413-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0414-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.