Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0414-2020Class II

Randox Liquid Cardiac Controls Catalogue Numbers CQ5053

Randox Laboratories, Limited / initiated 2019-10-04 / Terminated

Randox Laboratories, Limited began a recall of Randox Liquid Cardiac Controls Catalogue Numbers CQ5053 on . FDA classified it as Class II and gives the reason as "Removal of claims for N-proBNP in the Liquid Cardiac Control catalogue number CQ5053 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0414-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Randox Laboratories, Limited
Product
Randox Liquid Cardiac Controls Catalogue Numbers CQ5053.
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
Removal of claims for N-proBNP in the Liquid Cardiac Control catalogue number CQ5053 due to an observed decrease in recovery for N-Terminal Pro-Brain Natriuretic Peptide (NT-proBNP) in recent lots.
FDA root cause
Under Investigation by firm
Action
On October 1, 2019 the firm distributed Urgent Medical Device Recall letters by e-mail instructing them to: - Inspect your stock and quarantine affected stock. - Replace the value sheet in the kit with the revised value sheet provided. - Randox is not recommending a review of previous results as changes in quality control recovery would be reviewed at the time of occurrence. - Discuss the contents of this notice with your Medical Director. - Complete and return the response form 12187-QA to [email protected] within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization.
Quantity in commerce
6,702
Distribution
US Distribution: WV, CA, DE, NJ, NY, GA, IN, PR OUS Distribution: CA, PL, CH, JP, NL, IT, CN, GB, IE, RU, MD, BY, SA, VN, AE, BH, DK, CO, ES, FR, TW, PH, TH, OM, HU, IN, KR, ZA, AU, CZ, PE, CL, IR, NZ, BE, DE, BA, SV, SK, HK, PT, PK, BR, TR, UA, SE, QA, LT, PE,
Product codes and lots
Catalogue Number: CQ5053 GTIN: 05055273207460 Batch/lot: 4245CK (Exp. 28Nov2019), 4248CK (Exp. 28Nov2019), 4313CK (Exp. 28Jun2020) and 4316CK (Exp. 28Jun2020)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0414-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0412-2020Class IITerminatedVoluntary: Firm initiated
Z-0413-2020Class IITerminatedVoluntary: Firm initiated