Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0351-2020Class II

Lag Screw 3.2Mm Guide Pin Sleeve, Ref 71674032 -

Smith & Nephew, Inc. / initiated 2019-10-07 / Terminated / root cause: manufacturing process

Smith & Nephew, Inc. began a recall of LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft... on . FDA classified it as Class II and gives the reason as "LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully compress". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0351-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long oblique shaft fractures and segmental shaft fractures; subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; intracapsular fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; bone lengthening and shortening.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully compress.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm initiated the recall by letter and email on 10/07/2019. The letter explained the issue and requested the return of the recalled product.
Quantity in commerce
116 devices
Distribution
Worldwide distributions - US Nationwide distributions and countries of Belgium, Brazil, China, Columbia, France, Great Britain, India, Japan, Russia, Singapore, South Africa, Spain,and United Arab Emirates.
Product codes and lots
Batch Numbers: 18DM11613A, 18HM02575, 18HM13327A, 18HM20956A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0351-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0351-2020Class IITerminatedVoluntary: Firm initiated