Lag Screw 3.2Mm Guide Pin Sleeve, Ref 71674032 -
Smith & Nephew, Inc. began a recall of LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft... on . FDA classified it as Class II and gives the reason as "LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully compress". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0351-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Smith & Nephew, Inc.
- Product
- LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long oblique shaft fractures and segmental shaft fractures; subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; intracapsular fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; bone lengthening and shortening.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully compress.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm initiated the recall by letter and email on 10/07/2019. The letter explained the issue and requested the return of the recalled product.
- Quantity in commerce
- 116 devices
- Distribution
- Worldwide distributions - US Nationwide distributions and countries of Belgium, Brazil, China, Columbia, France, Great Britain, India, Japan, Russia, Singapore, South Africa, Spain,and United Arab Emirates.
- Product codes and lots
- Batch Numbers: 18DM11613A, 18HM02575, 18HM13327A, 18HM20956A
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0351-2020 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY