Giraffe Incubator Carestation, Giraffe Incubator Carestation CS1 infant warmer
GE Healthcare, LLC began a recall of Giraffe Incubator Carestation, Giraffe Incubator Carestation CS1 infant warmer on . FDA classified it as Class I and gives the reason as "Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0540-2020
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Giraffe Incubator Carestation, Giraffe Incubator Carestation CS1 infant warmer
- Product code
- FMT Warmer, Infant Radiant
- Reason for recall
- Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm disseminated its notices by letter beginning on 10/11/2019. The notices identified the following safety issues: - The bedside panels of Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation can be upright and look closed but not be latched. - The portholes also can look closed when not latched. - If a canopy cover is used, it can hold the bedside panel or porthole door closed without being latched. The product may continue to be used following specific safety instructions outlined in the notice. On/about 11/8/2019, the firm issued an "Amended - Urgent Medical Device Recall" letter that supplemented the previous notification to provide additional safety instructions and materials (poster regarding risk of patient fall, user manual addendum, and labeling, including safety labels). Update 6/3/2021: In the event a new consignee was identified, GE Healthcare has sent the original letter.
- Quantity in commerce
- 3467 units
- Distribution
- worldwide
- Product codes and lots
- Giraffe Incubator Carestation, Model Numbers:2082844-002-XXXXXX Updated 6/3/2021 - All model numbers and units are affected.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0538-2020 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0539-2020 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0540-2020 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0541-2020 | Class I | Terminated | Voluntary: Firm initiated |