Z-0540-2020Class I

Giraffe Incubator Carestation, Giraffe Incubator Carestation CS1 infant warmer

GE Healthcare / initiated 2019-10-11 / Terminated / root cause: design

GE Healthcare, LLC began a recall of Giraffe Incubator Carestation, Giraffe Incubator Carestation CS1 infant warmer on . FDA classified it as Class I and gives the reason as "Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0540-2020
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Giraffe Incubator Carestation, Giraffe Incubator Carestation CS1 infant warmer
Product code
FMT Warmer, Infant Radiant
Reason for recall
Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
Root cause tag
design
FDA root cause
Device Design
Action
The firm disseminated its notices by letter beginning on 10/11/2019. The notices identified the following safety issues: - The bedside panels of Giraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation can be upright and look closed but not be latched. - The portholes also can look closed when not latched. - If a canopy cover is used, it can hold the bedside panel or porthole door closed without being latched. The product may continue to be used following specific safety instructions outlined in the notice. On/about 11/8/2019, the firm issued an "Amended - Urgent Medical Device Recall" letter that supplemented the previous notification to provide additional safety instructions and materials (poster regarding risk of patient fall, user manual addendum, and labeling, including safety labels). Update 6/3/2021: In the event a new consignee was identified, GE Healthcare has sent the original letter.
Quantity in commerce
3467 units
Distribution
worldwide
Product codes and lots
Giraffe Incubator Carestation, Model Numbers:2082844-002-XXXXXX Updated 6/3/2021 - All model numbers and units are affected.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0540-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0538-2020Class ITerminatedVoluntary: Firm initiated
Z-0539-2020Class ITerminatedVoluntary: Firm initiated
Z-0540-2020Class ITerminatedVoluntary: Firm initiated
Z-0541-2020Class ITerminatedVoluntary: Firm initiated