Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0416-2020Class II

Guardia Access Embryo Transfer Catheter is used for transferring IVF embryos into the uterine cavity

Cook Incorporated / initiated 2019-10-03 / Terminated

Cook Inc. began a recall of Guardia Access Embryo Transfer Catheter is used for transferring IVF embryos into the uterine cavity on . FDA classified it as Class II and gives the reason as "Bent transfer catheter tips may cause difficulty or inability in inserting the transfer catheter into the guiding catheter. Potential adverse events that may occur if an affected product is used include increased...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0416-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cook Incorporated
Product
Guardia Access Embryo Transfer Catheter is used for transferring IVF embryos into the uterine cavity.
Product code
MQF Catheter, Assisted Reproduction
Reason for recall
Bent transfer catheter tips may cause difficulty or inability in inserting the transfer catheter into the guiding catheter. Potential adverse events that may occur if an affected product is used include increased procedural time, the need for a repeat embryo transfer procedure, or the need for a patient to undergo an additional IVF cycle.
FDA root cause
Under Investigation by firm
Action
An Urgent Field Safety Notice letter dated 10/3/19 was sent to customers. The letter instructs customers to identify affected devices, quarantine the devices, return them, and complete the enclosed Customer Response Form. Where product is indicated as being returned, our Customer Services department will contact you to organize the return and issue you with the relevant Returns Authorization number. Please include contact details on the Customer Response form. Returned Product should be addressed to: Cook Medical EUDC Robert-Koch-Strae, 2 52499 Baesweiler GERMANY Credit will be provided for the returned affected products where applicable.
Quantity in commerce
400
Distribution
No US distribution. The products were distributed to the following foreign countries: Belgium, Denmark, Germany, Italy, Netherlands, Norway, Poland, Republic of Macedonia, Slovenia, Spain, and Switzerland.
Product codes and lots
Lot Number 8361746

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0416-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0415-2020Class IITerminatedVoluntary: Firm initiated
Z-0416-2020Class IITerminatedVoluntary: Firm initiated