Guardia Access Embryo Transfer Catheter is used for transferring IVF embryos into the uterine cavity
Cook Inc. began a recall of Guardia Access Embryo Transfer Catheter is used for transferring IVF embryos into the uterine cavity on . FDA classified it as Class II and gives the reason as "Bent transfer catheter tips may cause difficulty or inability in inserting the transfer catheter into the guiding catheter. Potential adverse events that may occur if an affected product is used include increased...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0416-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cook Incorporated
- Product
- Guardia Access Embryo Transfer Catheter is used for transferring IVF embryos into the uterine cavity.
- Product code
- MQF Catheter, Assisted Reproduction
- Reason for recall
- Bent transfer catheter tips may cause difficulty or inability in inserting the transfer catheter into the guiding catheter. Potential adverse events that may occur if an affected product is used include increased procedural time, the need for a repeat embryo transfer procedure, or the need for a patient to undergo an additional IVF cycle.
- FDA root cause
- Under Investigation by firm
- Action
- An Urgent Field Safety Notice letter dated 10/3/19 was sent to customers. The letter instructs customers to identify affected devices, quarantine the devices, return them, and complete the enclosed Customer Response Form. Where product is indicated as being returned, our Customer Services department will contact you to organize the return and issue you with the relevant Returns Authorization number. Please include contact details on the Customer Response form. Returned Product should be addressed to: Cook Medical EUDC Robert-Koch-Strae, 2 52499 Baesweiler GERMANY Credit will be provided for the returned affected products where applicable.
- Quantity in commerce
- 400
- Distribution
- No US distribution. The products were distributed to the following foreign countries: Belgium, Denmark, Germany, Italy, Netherlands, Norway, Poland, Republic of Macedonia, Slovenia, Spain, and Switzerland.
- Product codes and lots
- Lot Number 8361746
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0415-2020 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0416-2020 | Class II | Terminated | Voluntary: Firm initiated |