Z-1154-2020Class II

Respironics RP-Touch Screen, V60/V680 Ventilator Serviced with Touchscreen RP Kit 453561511951 containing a...

Respironics, Inc. / initiated 2019-09-11 / Open, Classified

Respironics California, LLC began a recall of Respironics RP-Touch Screen, V60/V680 Ventilator Serviced with Touchscreen RP Kit 453561511951 containing a touchscreen labeled as raw part number 1132318... on . FDA classified it as Class II and gives the reason as "Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchcreen, and therefore the clinician...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1154-2020
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Respironics, Inc.
Product
Respironics RP-Touch Screen, V60/V680 Ventilator Serviced with Touchscreen RP Kit 453561511951 containing a touchscreen labeled as raw part number 1132318 Revision C
Product code
MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator
Reason for recall
Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchcreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia.
FDA root cause
Other
Action
On 09/26/19, Urgent Medical Device Correction Field Safety Notices were mailed to customers. Firm is asking customers to follow the Action to be Taken by Customer/User section of the Medical Device Correction notification/Field Safety Notice (Ventilator): 1) If in use on a patient, continue to use the ventilator if an alternative non-invasive ventilator is not available. 2) If a touchscreen failure is detected, disconnect the patient and immediately start ventilation with an alternate device. Contact your local customer service representative to report the failure. 3) To minimize the risk of patient injury, operate the Ventilator as defined in the operator's manual and promptly attend to all alarms presented by the ventilator. 4) As recommended in the operator's manual, use an external 02 monitor/analyzer and set the alarm thresholds appropriately. 5) As recommended in the operator's manual, have an alternative means of ventilation available whenever the ventilator is in use. 6) Acknowledge receipt of this notification. Firm is asking customers to follow the Action to be Taken by Customer/User section of the Medical Device Correction notification/Field Safety Notice (User Interface RP-Kits): 1) If the affected kit is installed on a ventilator and in use on a patient, continue to use the ventilator if an alternative non-invasive ventilator is not available. 2) If the affected kit is installed on a ventilator, document the serial number of the ventilator and the serial number of the affected User Interface Assembly it contained, on the attached Acknowledgement and Receipt Form. 3) If the affected devices could not be found or linked to a ventilator at your site, please indicate this within in the detail section provided on the form, 4) Return the completed and signed Acknowledgement and Receipt Form. Firm is asking customers to follow the Action to be Taken by Customer/User section of the Medical Device Correction notification/Field Safety Notice (Tou
Quantity in commerce
800
Distribution
US Nationwide. OUS (Foreign): Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Indonesia, Israel, Italy, Malaysia, Mexico, Netherlands, Portugal, Singapore, South Africa, Spain, Sweden, United Arab Emirates, United Kingdom, Viet Nam, Oman, Philippines, Saudi Arabia, Taiwan, Turkey, Russian Federation, Argentina, Ecuador, Poland, Guatemala, Peru, Viet Nam
Product codes and lots
Touchscreen Serial Number: 41088-1808-0001 to 0179,  41213-1809-0001 to 0200,  41305-1811-0001 to 0295,  41379-1812-0001 to 0250,  41379-1812-0251 to 0328,  41540-1813-0001 to 0195,  41633-1814-0001 to 0275,  41633-1814-0276 to 0285,  41724-1815-0001 to 0140,  41724-1815-0141 to 0183,  41838-1816-0001 to 0107,  41838-1816-0108 to 0165,  41930-1817-0001 to 0092,  42059-1818-0001 to 0150,  42059-1818-0151 to 0183,  42161-1819-0001 to 0117,  42161-1819-0118 to 0137,  42232-1820-0001 to 0130,  42232-1820-0131 to 0146,  42334-1822-0001 to 0134,  42334-1822-0135 to 0141,  42471-1823-0001 to 0143,  42471-1823-0144 to 0175,  42805-1828-0001 to 0138

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1154-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1152-2020Class IIOngoingVoluntary: Firm initiated
Z-1153-2020Class IIOngoingVoluntary: Firm initiated
Z-1154-2020Class IIOngoingVoluntary: Firm initiated
Z-1155-2020Class IIOngoingVoluntary: Firm initiated