Software versions syngo CT VB20 running on the following Siemens SOMATOM CT Scanner. SOMATOM Force (Model...
Siemens Medical Solutions USA, Inc began a recall of Software versions syngo CT VB20 running on the following Siemens SOMATOM CT Scanner. SOMATOM Force (Model #10742326), SOMATOM Definition As (Model #8098027)... on . FDA classified it as Class II and gives the reason as "Software issue identified in the software versions syngo CT VB20 running on the SOMATOM CT Scanner: scan aborts and system crashes". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0406-2020
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Software versions syngo CT VB20 running on the following Siemens SOMATOM CT Scanner. SOMATOM Force (Model #10742326), SOMATOM Definition As (Model #8098027), SOMATOM Definition Edge (Model #10590000), SOMATOM Definition Flash (Model #10430603), SOMATOM Drive (Model #10431700), SOMATOM Confidence (Model #10590100), and SOMATOM Edge Plus (Model # 10267000) with software syngo CT VB20
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Software issue identified in the software versions syngo CT VB20 running on the SOMATOM CT Scanner: scan aborts and system crashes
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens Medical Solutions USA, Inc. issued a Customer Safety Advisory Notice (CSAN) to all affected customers via CT070/19/S. on 10/719 via email/or letter. The CSAN provides recommendations to avoid the issues until the technical solution can be applied. The issues will be resolved with software update version syngo CT VB20_SP1 (Service Pack 1). You will receive the updated software free of charge. Questions or require technical support, please contact the service organization at 1-800-888-7436.
- Quantity in commerce
- 305 units
- Distribution
- Nationwide
- Product codes and lots
- Software versions syngo CT VB20
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0406-2020 | Class II | Terminated | Voluntary: Firm initiated |