NIDEK Specular Microscope, Model CEM-530 -
Nidek, Inc. began a recall of NIDEK Specular Microscope, Model CEM-530 - Product Usage: The device provides noncontact, high magnification image capture of endothelium enabling observation... on . FDA classified it as Class III and gives the reason as "The electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0771-2020
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Nidek Incorporated
- Product
- NIDEK Specular Microscope, Model CEM-530 - Product Usage: The device provides noncontact, high magnification image capture of endothelium enabling observation of the size and shape of cells. Information such as the number of endothelial cells, cell density, and cell area is analyzed through the captured images.
- Product code
- NQE Microscope, Specular
- Reason for recall
- The electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The recalling firm issued letters dated 2/7/2019 via FedEx explaining the issue, informing the consignee there was no risk associated with use of the device, and that the firm is preparing for the replacement parts in order to schedule for the correction to be made.
- Quantity in commerce
- 13 devices
- Distribution
- US Nationwide distribution in the states of AZ, CA, IL, KS, MD, NJ, NY, PA, SC, and VA, as well as PR.
- Product codes and lots
- Serial numbers 120080, 120081, 120086, 120087, 120088, 120090, 120091, 120092, 120093, 120094, 120095, 120096, and 170001.