Z-0498-2020Class II

HeartStart XL+ Defibrillator/Monitor, Model 861290

Philips North America LLC / initiated 2019-10-16 / Terminated / root cause: manufacturing process

Philips North America, LLC began a recall of HeartStart XL+ Defibrillator/Monitor, Model 861290 on . FDA classified it as Class II and gives the reason as "The rotary therapy selector switch may fail, resulting in unexpected device behavior. These behaviors include: The device may not turn on; The device may not perform the selected function; The device may deliver a shock...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0498-2020
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
HeartStart XL+ Defibrillator/Monitor, Model 861290
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
The rotary therapy selector switch may fail, resulting in unexpected device behavior. These behaviors include: The device may not turn on; The device may not perform the selected function; The device may deliver a shock with an energy level different from the setting selected by the user.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On October 16, 2019, Philips Healthcare issued Urgent Medical Device Correction notices to U.S. customers via certified mail. Philips Markets Organizations are responsible for distributing the letters outside of the U.S. Actions to be taken by the customer: 1) The device is safe to use and can remain in service if it does not exhibit any of these behaviors described in this Notice. 2) Continue to perform Shift Checks and Operational checks as recommended in the Instructions for Use (IFU) as this reduces the risk of a failure during use. 3) If you identify a device that exhibit any of these behaviors; please remove it from service and contact Philips to request service. Philips will contact you to arrange for repair of your unit once parts are available. Philips will install a replacement switch in affected devices at no charge to the customer. If you need further information or support concerning this notification, please contact your local Philips representative.
Quantity in commerce
23,749 Units
Distribution
US Nationwide
Product codes and lots
All units manufactured prior to 1 May 2017.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0498-2020
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0498-2020Class IITerminatedVoluntary: Firm initiated